Quality Control & Error Analysis: Mixed Analytical Practice
Quality control and error analysis dominate; hardness and metal-analysis items are a small strand.
Assessment Overview: 5 multiple-choice questions assessing conceptual and mathematical proficiency in Quality Control & Error Analysis: Mixed Analytical Practice.
Practice Questions & Answers
Question 1: Each link in the traceability chain adds
- A. accuracy
- B. nothing
- C. uncertainty (Correct Answer)
- D. money only
Pedagogical Explanation: Uncertainties propagate down the chain; the end user inherits them all.
Question 2: Error scenario: Small timing differences when starting a stopwatch The error type here is:
- A. Not an error at all
- B. Systematic error
- C. Random error (Correct Answer)
- D. A gross mistake
Pedagogical Explanation: It scatters measurements unpredictably (affects precision); reduced by replication and statistics.
Question 3: Error scenario: A balance that reads 0.05 g high on every weighing How should this error be classified?
- A. A gross mistake
- B. Systematic error (Correct Answer)
- C. Not an error at all
- D. Random error
Pedagogical Explanation: It biases every measurement in the same direction (affects accuracy); found by blanks, CRMs and calibration.
Question 4: Finding: electronic data with lost passwords. The ALCOA principle most directly violated is:
- A. Complete
- B. Consistent
- C. Available (Correct Answer)
- D. Original
Pedagogical Explanation: electronic data with lost passwords violates Available (retrievable for review/audit).
Question 5: QC description: pure reagents run to isolate their contribution. This is a:
- A. trip blank
- B. method blank
- C. reagent blank (Correct Answer)
- D. calibration blank