Analytical Statistics, Calibration & Quality Practice
Analytical statistics, calibration curves, quality control and detection limits — a general analytical review.
Assessment Overview: 5 multiple-choice questions assessing conceptual and mathematical proficiency in Analytical Statistics, Calibration & Quality Practice.
Practice Questions & Answers
Question 1: Linearity study r = 0.99089. Against r >= 0.999, this is:
- A. perfect
- B. irrelevant
- C. acceptable
- D. unacceptable (Correct Answer)
Pedagogical Explanation: r = 0.99089 fails the >= 0.999 criterion.
Question 2: TISAB also contains complexing agents to
- A. add fluoride
- B. precipitate fluoride
- C. free F⁻ from Al³⁺ and Fe³⁺ complexes (Correct Answer)
- D. change the color
Pedagogical Explanation: Agents like citrate or CDTA decomplex aluminum- and iron-bound fluoride so total F⁻ is measured.
Question 3: Six replicate assays: 100.2, 100.2, 100, 100.3, 100.2, 100.5. Repeatability RSD =
- A. 0.33%
- B. 0.163%
- C. 0.16% (Correct Answer)
- D. 0.08%
Pedagogical Explanation: RSD = s/mean x 100% = 0.163/100.23 x 100% = 0.16%.
Question 4: Calibration practice: balance calibrated with certified masses linked to the national standard. This practice is:
- A. mandatory
- B. irrelevant
- C. traceable practice (Correct Answer)
- D. not traceable
Pedagogical Explanation: An unbroken documented chain to SI references exists.
Question 5: Gravimetric factor for magnesium from Mg2P2O7 is 2M(Mg)/M(Mg2P2O7) = 0.2184. A 0.58 g precipitate contains how many g of analyte?
- A. 0.2534
- B. 0.1267 (Correct Answer)
- C. 0.0634
- D. 2.6551
Pedagogical Explanation: Mass analyte = mass ppt × GF = 0.58 × 0.2184 = 0.1267 g.