Mixed Analytical Chemistry Practice: Chromatography & Quality Control
A mixed analytical set: chromatography and quality-control questions with a small carbonyl strand.
Assessment Overview: 5 multiple-choice questions assessing conceptual and mathematical proficiency in Mixed Analytical Chemistry Practice: Chromatography & Quality Control.
Practice Questions & Answers
Question 1: Spiking should be done
- A. into the blank
- B. into pure water only
- C. into the real sample matrix before preparation (Correct Answer)
- D. after analysis
Pedagogical Explanation: Matrix spikes reveal losses and matrix effects through the full procedure.
Question 2: A 3.3 ppm standard gives S/N = 9. Estimated LOD (S/N = 3) is:
- A. 1.1 ppm (Correct Answer)
- B. 0.37 ppm
- C. 3.3 ppm
- D. 3.67 ppm
Pedagogical Explanation: LOD = conc x 3/(S/N) = 3.3 x 3/9 = 1.1 ppm.
Question 3: QC description: a blank that rides along during the whole field campaign. This is a:
- A. method blank
- B. field blank (Correct Answer)
- C. trip blank
- D. equipment blank
Pedagogical Explanation: 'a blank that rides along during the whole field campaign' describes the field blank.
Question 4: A common cause of peak tailing in LC is
- A. too high a flow rate
- B. too small an injection
- C. interaction of basic analytes with residual silanol groups (Correct Answer)
- D. detector overload
Pedagogical Explanation: Basic compounds adsorb strongly to acidic residual silanols on silica-based columns, producing tailing.
Question 5: The most commonly used GC carrier gas is
- A. oxygen
- B. helium (Correct Answer)
- C. ammonia
- D. carbon dioxide
Pedagogical Explanation: Helium is the most popular carrier: inert, safe, and efficient over a wide velocity range.