Operational Principle

Identifies and quantifies elements in solids, liquids and powders by measuring the characteristic X-ray fluorescence emitted when atoms are excited by a primary X-ray beam. It is non-destructive and covers elements from sodium (Z=11) to uranium (Z=92).

Optical & Mechanical Architecture

A rhodium or tungsten X-ray tube (typically 4–50 kV) excites inner-shell electrons; when outer electrons fill the vacancy, element-specific Kα, Kβ and Lα lines are emitted. In wavelength-dispersive systems a crystal monochromator (Bragg diffraction, nλ = 2d·sinθ) resolves lines down to ~0.005 nm; in energy-dispersive systems a silicon drift detector sorts photons by energy. Moseley's law relates line energy to atomic number, so quantification is matrix-matched to fundamental parameters or certified standards.

Applications in Research & Industry

Alloy grade identification and scrap sorting in metals industry,Cement, mining and soil elemental mapping,RoHS screening for lead, cadmium and mercury in electronics,Pigment and glaze analysis in art conservation

Typical Performance Specifications

Element range Na–U (EDXRF), ppm to 100% · tube 4–50 kV · typical LOD 1–10 ppm for transition metals