UV-Vis Spectrophotometer
Measures how much light a sample absorbs
Operational Principle
Shines ultraviolet and visible light through a sample and measures absorbance at each wavelength — the basis of colorimetry and concentration analysis.
Optical & Mechanical Architecture
A deuterium + tungsten lamp covers 190–1100 nm; a monochromator selects one wavelength; the beam splits into sample and reference paths; detectors compare intensities. Absorbance follows the Beer–Lambert law A = εbc.
Applications in Research & Industry
Concentration assays,Kinetics (absorbance vs time),Water quality (nitrate, phosphate),DNA purity (A260/A280)
Typical Performance Specifications
Range 190–1100 nm · bandwidth 1–2 nm · photometric accuracy ±0.003 AU
Operational Visualization Notes
spectrometer