Raman Optical Activity Spectrometer (ROA)
Chiral fingerprints in scattered Raman light
Operational Principle
A Raman optical activity spectrometer that detects the minute intensity difference in Raman scattering for left- versus right-circularly polarized incident light — the Raman analogue of circular dichroism.
Optical & Mechanical Architecture
In backscattering geometry a laser beam of alternating circular polarization excites the sample and the scattered Raman light is collected at 180 degrees; the tiny differential (often 10^-3 to 10^-5) builds the ROA spectrum. It is especially sensitive to protein and carbohydrate three-dimensional structure in water, where VCD struggles.
Applications in Research & Industry
Protein secondary and tertiary structure in solution,Carbohydrate conformation and anomers,Absolute configuration without crystals,Chiral solvation and hydration studies
Typical Performance Specifications
Operational Visualization Notes
spectrometer