Operational Principle

Measures phosphorescence — the delayed emission from triplet states — to analyze compounds that fluoresce weakly or not at all. Its time-gated detection separates long-lived phosphorescence from prompt fluorescence and scatter.

Optical & Mechanical Architecture

A xenon flash lamp excites the sample; a rotating-can or electronic chopper opens the detector only after a delay of ~0.1–10 ms, when fluorescence and scatter have decayed. The remaining phosphorescence (lifetimes ms to seconds, obeying first-order decay I = I₀e^(−t/τ)) is recorded as excitation/emission spectra or decay curves. Samples are often frozen at 77 K or immobilized on filter paper to protect triplet states from oxygen quenching.

Applications in Research & Industry

Polycyclic aromatic hydrocarbon (PAH) analysis,Drug photostability and impurity profiling,Oxygen sensing via phosphorescence quenching,Forensic ink and document examination

Typical Performance Specifications

Delay 0.1–10 ms gate · lifetime range µs–s · 77 K cryogenic option · sensitivity sub-ppb for PAHs