Photoacoustic Spectrometer

Hears light absorption in opaque solids and gases

SPECTROSCOPYAnalytical Instrument

Operational Principle

Detects absorbed light via the tiny pressure wave (sound) it generates, so it works on opaque solids, powders and turbid samples where ordinary transmission spectroscopy fails. The acoustic signal is directly proportional to absorbed energy.

Optical & Mechanical Architecture

Modulated monochromatic light (chopped at audio frequencies, ~20–500 Hz) strikes the sample in a sealed cell; absorbed energy thermalizes and launches a pressure wave detected by a sensitive microphone — the Rosencwaig–Gersho effect. Scanning wavelength produces a spectrum; varying modulation frequency changes the thermal diffusion length (µ = √(2α/ω)), giving depth-resolved profiles of layered materials like coatings and skin.

Applications in Research & Industry

Trace gas sensing (CO, CH₄, NOx) at ppb levels,Pigment layer analysis in paintings,Blood glucose and tissue oxygenation research,Catalyst surface absorption studies

Typical Performance Specifications

UV to mid-IR sources · modulation 20 Hz–5 kHz · gas LOD down to ppb · depth profiling 1–500 µm

Operational Visualization Notes

spectrometer