Magnetic Force Microscope (MFM)
Nanoscale magnetic domain imaging with a tip
Operational Principle
The magnetic force microscope maps magnetic domains and stray fields at ~50 nm resolution using a magnetized AFM tip scanned in lift mode above the surface.
Optical & Mechanical Architecture
A cantilever coated with a ferromagnetic film (CoCr, ~30–50 nm) is first scanned in tapping mode to record topography; it then retraces the same line lifted 20–100 nm above the surface, where only long-range magnetic forces act. The tip–sample magnetic interaction shifts the cantilever's resonant frequency or phase, producing a map of the stray-field gradient — domain walls and bits appear as contrast. Two-pass lift mode cleanly separates magnetic from topographic signals.
Applications in Research & Industry
Hard-disk bit and domain-wall imaging,Magnetic nanoparticle and thin-film domain studies,Spintronic device and MRAM characterization,Superconducting vortex lattice imaging
Typical Performance Specifications
Operational Visualization Notes
microscope