Modulated DSC (mDSC)
Separates overlapping thermal events
Operational Principle
Overlays a small sinusoidal temperature modulation on the DSC ramp, splitting the heat flow into reversing (heat capacity) and non-reversing (kinetic) components.
Optical & Mechanical Architecture
While ramping at ~2 °C/min, the temperature oscillates ±0.5–1 °C every 40–100 s. The in-phase response gives reversible Cp (glass transitions); the out-of-phase part captures crystallization, curing, and relaxation that ordinary DSC smears together.
Applications in Research & Industry
Glass transition in curing resins,Weak Tg detection,Crystallization vs melting overlap,Pharma amorphous content
Typical Performance Specifications
±0.5–1 °C modulation · reversing/non-reversing split · 40–100 s period
Operational Visualization Notes
thermal