SPECTROSCOPYAnalytical Instrument

Operational Principle

Fires a pulsed laser at a material to create a micro-plasma and reads its emission lines for elemental composition — no sample preparation, works on solids in air, and can detect light elements like hydrogen, lithium and carbon that XRF misses.

Optical & Mechanical Architecture

A Q-switched Nd:YAG laser (1064 nm, ~100–200 mJ per pulse, 5–10 ns) ablates a ~100 µm spot, heating it to 10,000–20,000 K. As the plasma cools, excited atoms emit characteristic lines recorded by a CCD spectrometer with gated delay (0.5–2 µs) to avoid the continuum flash. Calibration curves or calibration-free plasma models (assuming local thermal equilibrium, Saha–Boltzmann) convert line intensities to concentrations.

Applications in Research & Industry

Alloy verification and positive material identification (PMI),Coal quality and ash-content analysis,Archaeological artifact provenance studies,Planetary exploration (e.g., rover-mounted LIBS on Mars)

Typical Performance Specifications

Laser 1064 nm, ~100 mJ/pulse · spot ~100 µm · elements H–U · LOD 1–100 ppm · single-shot ~1 s

Operational Visualization Notes

beam