CHEMVEN

Four-Point Probe Resistivity Tester

Sheet resistance without contact errors

ELECTROCHEMISTRYAnalytical Instrument

Operational Principle

Measures the electrical resistivity of thin films and semiconductors with four collinear probes, eliminating the contact resistance that ruins two-probe measurements.

Optical & Mechanical Architecture

Current is forced through the outer two probes while the inner two sense voltage with a high-impedance meter, so probe-contact resistance drops out of the reading. For a thin film, sheet resistance R_s = (π/ln 2)·(V/I); thickness gives bulk resistivity.

Applications in Research & Industry

Thin-film QC,Silicon wafer resistivity,ITO/transparent conductors,Battery electrode coatings

Typical Performance Specifications

4 collinear probes · 1 mA–100 mA · sheet Ω/□ and bulk Ω·cm

Operational Visualization Notes

electrochem

← Back to Instruments Hub