Environmental SEM (ESEM)
SEM of wet and insulating samples without coating
Operational Principle
The environmental SEM images hydrated, oily or insulating samples at low vacuum, using gas ionization to neutralize charge and amplify the signal — no coating or drying needed.
Optical & Mechanical Architecture
The specimen chamber holds a controlled gas (usually water vapor, 1–2000 Pa) while the electron column stays at high vacuum via differential pumping apertures. Primary electrons scatter off gas molecules, and secondary electrons ionize the gas in a cascade that amplifies the signal (gaseous secondary electron detection); positive ions drift back to the sample, neutralizing surface charge on insulators. A Peltier stage controls temperature so humidity can be held at saturation, keeping biological samples hydrated during imaging.
Applications in Research & Industry
Imaging wet biological tissue, plants and food without drying,Uncoated polymers, ceramics and geological samples,In situ hydration/dehydration and corrosion studies,Dynamic wetting and crystallization observation
Typical Performance Specifications
Operational Visualization Notes
microscope