EPR / ESR Spectrometer
Detects unpaired electrons and free radicals
Operational Principle
Detects species with unpaired electrons — free radicals, transition-metal complexes and defect centers — by their absorption of microwaves in a magnetic field. It is the only technique that sees radicals directly, down to nanomolar spin concentrations.
Optical & Mechanical Architecture
The sample sits in a microwave cavity (usually X-band, ~9.8 GHz, with fields swept 0–600 mT around resonance g·μB·B = hν). When the Zeeman splitting of the electron spin matches the photon energy, absorption is recorded as a first-derivative signal. The g-factor reveals the chemical environment; hyperfine splitting from neighboring nuclei identifies the radical's structure. Spin traps (e.g., DMPO) capture short-lived radicals for indirect detection.
Applications in Research & Industry
Antioxidant and radical-scavenging capacity assays,Defect and F-center studies in semiconductors and catalysts,Radiation dosimetry via tooth enamel or alanine,Metalloprotein active-site characterization
Typical Performance Specifications