CHEMVEN

Electrolyzer Test Rig

Water-splitting performance and gas analysis

ELECTROCHEMISTRYAnalytical Instrument

Operational Principle

A test rig for PEM or alkaline water electrolyzers that drives water splitting at controlled current density, measures cell voltage efficiency, and quantifies evolved H₂/O₂ with gas analysis.

Optical & Mechanical Architecture

Deionized water is circulated to the stack while a DC supply applies current; evolved gases are separated, dried, and metered (mass flow or GC) to compute faradaic efficiency. Temperature, pressure, and water conductivity are logged throughout.

Applications in Research & Industry

Catalyst and membrane screening for green hydrogen,Efficiency vs current-density mapping of stacks,Degradation and crossover measurements in PEM cells

Typical Performance Specifications

Current density up to 6 A/cm²; cell temp 25–90 °C; pressure up to 30 bar; faradaic efficiency via H₂ mass flow ±1%

Operational Visualization Notes

electrochem

← Back to Instruments Hub