EDS / EDX Microanalysis
Elemental analysis from characteristic X-rays
Operational Principle
Energy-dispersive X-ray spectroscopy identifies and quantifies elements in the electron microscope by measuring the energies of characteristic X-rays emitted under electron bombardment.
Optical & Mechanical Architecture
The electron beam ejects inner-shell electrons; outer electrons drop into the vacancies, emitting X-rays at element-specific energies (Moseley's law) — e.g. Fe Kα at 6.40 keV, O Kα at 0.525 keV. A silicon drift detector (SDD) measures each photon's energy, building a spectrum from ~0.1 keV upward; peak intensities are converted to composition via ZAF or φ(ρz) matrix corrections. Modern SDDs collect 10^5–10^6 counts/s, enabling fast elemental maps; detection covers roughly Be (Z=4) to U (Z=92).
Applications in Research & Industry
Elemental mapping of alloys, minerals and devices,Particle, inclusion and contaminant identification,Coating thickness and multilayer composition checks,Corrosion product and failure-debris analysis
Typical Performance Specifications