Clark Oxygen Electrode
Amperometric dissolved-oxygen sensing
Operational Principle
A membrane-covered amperometric electrode that measures dissolved oxygen concentration via its reduction current at a platinum cathode, widely used in environmental analysis and respirometry.
Optical & Mechanical Architecture
O₂ diffuses through a gas-permeable membrane into an internal electrolyte and is reduced at the Pt cathode (held at ~−0.6 V vs Ag/AgCl); the resulting current is proportional to the oxygen partial pressure outside the membrane. Temperature compensation corrects for solubility changes.
Applications in Research & Industry
Dissolved-oxygen monitoring in water and wastewater,Cellular and mitochondrial respiration assays,Fermentation and aquaculture process control
Typical Performance Specifications
Range 0–20 mg/L O₂ (0–200% saturation); response time < 30 s; accuracy ±1% FS; temp compensation 0–50 °C
Operational Visualization Notes
electrochem