Conductive AFM (C-AFM)
Nanoscale conductivity mapping with a biased probe
Operational Principle
An atomic force microscope variant with a conductive tip that simultaneously records topography and local electrical current, exposing conductive pathways, defects, leakage sites and resistive switching at nanometer resolution.
Optical & Mechanical Architecture
A metal-coated or solid platinum tip scans in contact mode while a bias is applied between tip and sample. The resulting current (pA to μA) is amplified and plotted alongside topography; I-V curves can be recorded at individual points by ramping the bias with the tip held stationary.
Applications in Research & Industry
Locating leakage paths and breakdown spots in thin dielectrics,Mapping conductive filaments in resistive RAM devices,Studying grain-boundary conductivity in polycrystalline oxides,Characterizing local conductivity of 2D materials and nanowires
Typical Performance Specifications
Operational Visualization Notes
microscope