Conductive AFM (C-AFM)

Nanoscale conductivity mapping with a biased probe

MICROSCOPYAnalytical Instrument

Operational Principle

An atomic force microscope variant with a conductive tip that simultaneously records topography and local electrical current, exposing conductive pathways, defects, leakage sites and resistive switching at nanometer resolution.

Optical & Mechanical Architecture

A metal-coated or solid platinum tip scans in contact mode while a bias is applied between tip and sample. The resulting current (pA to μA) is amplified and plotted alongside topography; I-V curves can be recorded at individual points by ramping the bias with the tip held stationary.

Applications in Research & Industry

Locating leakage paths and breakdown spots in thin dielectrics,Mapping conductive filaments in resistive RAM devices,Studying grain-boundary conductivity in polycrystalline oxides,Characterizing local conductivity of 2D materials and nanowires

Typical Performance Specifications

Current range ~1 pA-10 μA; spatial resolution ~10-30 nm; bias ±10 V; compatible with temperature and illumination stages