CHEMVEN

Accelerating Rate Calorimeter (ARC)

Adiabatic runaway-reaction safety testing

THERMAL ANALYSISAnalytical Instrument

Operational Principle

An adiabatic calorimeter that detects the onset of exothermic decomposition and tracks self-heating rate and pressure during thermal runaway — the standard tool for chemical process safety assessment.

Optical & Mechanical Architecture

The sample bomb is heated in heat–wait–search steps inside an adiabatic shield that tracks the sample temperature. Once exotherm onset is found, the system follows the runaway adiabatically, recording dT/dt and pressure vs T to derive TMR (time to maximum rate) and vent-sizing data.

Applications in Research & Industry

Thermal-hazard screening of reactive chemicals (onset, TMR),Battery cell thermal-runaway characterization,Relief vent sizing per DIERS methodology

Typical Performance Specifications

Temp up to 500 °C; pressure up to 200 bar; self-heat rate detection ~0.02 °C/min; sample 1–10 g; phi-factor correction

Operational Visualization Notes

furnace

← Back to Instruments Hub