CHEMVEN

Heat Capacity of the Calorimeter by Electrical Calibration

To determine the total heat capacity of a calorimeter by electrical heating (V·I·t method).

Level: UniversitySafety: Needs supervisionTime: 75 min
Aim of Experiment: To determine the total heat capacity of a calorimeter by electrical heating (V·I·t method).

Theoretical Foundations

Passing current I at voltage V for time t dissipates q = VIt joules into the calorimeter. With water of known mass present, (m_w·c_w + C_cal)ΔT = VIt, giving C_cal directly. Electrical calibration is the most accurate route to the calorimeter constant because the energy input is measured electrically, not thermally.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Calorimeter with immersion heater
  • DC power supply
  • Voltmeter
  • Ammeter
  • Stopwatch
  • Thermometer (0.1 °C)
  • Balance

Chemical Reagents

  • Distilled water

Step-by-Step Practical Procedure

  1. Place 200.0 g water in the calorimeter; record T.
  2. Set ~6.0 V across the heater; record V and I.
  3. Switch on for exactly 120 s with stirring.
  4. Record the temperature rise ΔT.
  5. C_cal = (VIt/ΔT) - m_w·c_w.
  6. Use this C_cal to correct a neutralisation result.

Observations & Expected Results

Typical student data: V = 6.0 V, I = 2.0 A, t = 120 s, ΔT = 1.60 °C with 200.0 g water → C_cal ≈ 64 J/°C for a glass calorimeter (larger than a polystyrene cup).

Worked Stoichiometric Calculations

Problem: Typical student data: V = 6.0 V, I = 2.0 A, t = 120 s, 200.0 g water, ΔT = 1.60 °C. Find C_cal.

Working: q = VIt = 6.0 × 2.0 × 120 = 1440 J. Total heat capacity = 1440/1.60 = 900 J/°C. Water contributes 200.0 × 4.18 = 836 J/°C, so C_cal = 900 - 836 = 64 J/°C. (Sanity check: C_total must exceed the water contribution — a useful trap for sign/unit errors.)

Safety Precautions & Hazard Control

  • Supervision: electrical heating near water — check insulation and use low voltage.
  • Do not run the heater dry.
  • Stir to avoid local boiling at the element.

Waste Disposal Protocol

Water to drain.

Viva Voce Oral Defense Questions & Answers

Q: Why is electrical calibration more accurate than the mixing method?

A: Electrical energy VIt is measured with high precision; no hot-water transfer losses occur.

Q: What does C_cal include?

A: The heat capacity of the vessel, lid, stirrer, thermometer and heater — everything that warms with the water.

Q: The worked example first gave an impossible result — what was wrong?

A: The stated ΔT was inconsistent with the energy input; checking that q/ΔT exceeds the water's own heat capacity is an essential sanity test.

← Back to Experiments Directory