Surface Tension by the Du Nouy Ring Tensiometer
To measure surface tension with a Du Nouy ring and compare with stalagmometer values.
Theoretical Foundations
A platinum ring pulled from the liquid surface detaches at force F = 2·(2·pi·R)·gamma (two sides of the film). Measuring F with a torsion balance gives gamma directly - the standard laboratory method.
Required Apparatus & Chemicals
Laboratory Apparatus
- Du Nouy tensiometer (or improvised ring + balance)
- Platinum-iridium ring
- Beakers
- Thermometer
Chemical Reagents
- Distilled water
- Ethanol
- SDS solution (optional)
Step-by-Step Practical Procedure
- Clean the ring by flaming.
- Immerse it in water; zero the balance.
- Raise the platform slowly until the ring detaches; record F.
- Calculate gamma = F/(4·pi·R) with correction factor.
- Repeat for ethanol and compare with drop-count results.
Observations & Expected Results
Typical student data: water 72.5 mN/m, ethanol 22.8 mN/m - agreeing with stalagmometer values within 2%.
Worked Stoichiometric Calculations
Problem: Ring radius R = 1.00 cm; detachment force F = 9.10 x 10^-3 N. Find gamma.
Working: gamma = F/(4·pi·R) = 9.10 x 10^-3/(4 x 3.1416 x 0.0100) = 9.10 x 10^-3/0.12566 = 0.0724 N/m = 72.4 mN/m.
Safety Precautions & Hazard Control
- Flame the ring before each measurement.
- Pull slowly and vertically.
- The liquid surface must be clean and still.
Waste Disposal Protocol
Test liquids to appropriate waste.
Viva Voce Oral Defense Questions & Answers
Q: Why the factor 4·pi·R?
A: The film has two surfaces (inner and outer), each of circumference 2·pi·R.
Q: Why flame the platinum ring?
A: To burn off organic contamination that would lower the measured force.
Q: When is a correction factor needed?
A: For precise work: the ring lifts extra liquid weight (Harkins-Jordan correction).