Potassium Ferrioxalate Actinometer: Measuring Light Intensity

To determine the photon flux of a lamp using the potassium ferrioxalate chemical actinometer.

Level: UniversitySafety: Needs supervisionTime: 2 h
Aim of Experiment: To determine the photon flux of a lamp using the potassium ferrioxalate chemical actinometer.

Theoretical Foundations

Ferrioxalate photoreduces with known quantum yield (Phi = 1.21 at 436 nm): 2[Fe(C2O4)3]^3- + hv -> 2[Fe(C2O4)2]^2- + 2CO2 + C2O4^2-. The Fe^2+ formed is complexed with 1,10-phenanthroline (red, 510 nm); from its moles and Phi the photon flux (einstein/s) follows.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Photolysis cell
  • Lamp with 436 nm filter
  • Spectrophotometer/colorimeter
  • Volumetric flasks
  • Stopwatch

Chemical Reagents

  • Potassium ferrioxalate (0.006 M in 0.05 M H2SO4)
  • 1,10-phenanthroline solution
  • Sodium acetate buffer
  • Sulfuric acid

Step-by-Step Practical Procedure

  1. Fill the cell with ferrioxalate solution; irradiate for a measured time (e.g. 10 min).
  2. Mix an aliquot with phenanthroline + buffer; develop the red complex.
  3. Measure absorbance at 510 nm; use epsilon = 11,100 L/(mol·cm) to find [Fe^2+].
  4. Moles Fe^2+ = c·V; photons absorbed = moles/Phi.
  5. Divide by irradiation time for flux in einstein/s.

Observations & Expected Results

The irradiated solution develops a strong red colour with phenanthroline; a 10 min exposure typically gives A_510 ~ 0.5-1.0.

Worked Stoichiometric Calculations

Problem: After 600 s irradiation, 25 mL of photolysed solution (diluted to 50 mL with reagents) shows A = 0.670 at 510 nm (1 cm cell, epsilon = 11100). Phi = 1.21. Find the photon flux.

Working: c in measured solution = 0.670/11100 = 6.04 x 10^-5 M. Moles Fe^2+ in 50 mL = 6.04 x 10^-5 x 0.050 = 3.02 x 10^-6 mol (from 25 mL photolysed). Photons = 3.02 x 10^-6/1.21 = 2.50 x 10^-6 einstein. Flux = 2.50 x 10^-6/600 = 4.2 x 10^-9 einstein/s.

Safety Precautions & Hazard Control

  • Prepare ferrioxalate fresh and keep it dark before use.
  • Wear gloves - oxalates are toxic.
  • Handle the lamp carefully (heat/UV).

Waste Disposal Protocol

Iron-oxalate waste to heavy-metal/organic waste as per local rules.

Viva Voce Oral Defense Questions & Answers

Q: What is a chemical actinometer?

A: A system with known quantum yield used to measure photon flux by chemical analysis.

Q: Why is ferrioxalate the standard?

A: High quantum yield (~1.2), wavelength-independent response 250-500 nm, and easy Fe^2+ analysis.

Q: What is an einstein?

A: One mole of photons (6.022 x 10^23 photons).