Potentiometric Argentometric Titration: Chloride with AgNO3
To determine chloride by potentiometric titration with AgNO3 using a silver electrode.
Theoretical Foundations
Ag+ + Cl- -> AgCl(s). A silver indicator electrode follows E = Edeg + 0.0591 log[Ag+]. Before the endpoint [Ag+] is tiny (fixed by Ksp); after, it rises sharply, giving a large potential jump at the endpoint located from DeltaE/DeltaV.
Required Apparatus & Chemicals
Laboratory Apparatus
- Silver wire indicator electrode
- Calomel reference electrode
- Digital voltmeter
- Burette
- Beaker
- Stirrer
Chemical Reagents
- Sodium chloride (unknown, ~0.05 M)
- Silver nitrate (0.050 M standard)
- Distilled water (for rinsing glassware and electrodes)
Step-by-Step Practical Procedure
- Pipette 25.0 mL of the chloride sample.
- Insert the Ag and reference electrodes.
- Titrate with 0.050 M AgNO3, recording E in 0.5 mL steps (0.1 mL near the jump).
- Plot E and DeltaE/DeltaV vs V.
- Find the endpoint and calculate [Cl-].
Observations & Expected Results
E ~ 0.3 V initially, jumping ~200 mV at the endpoint.
Worked Stoichiometric Calculations
Problem: Endpoint at 24.4 mL of 0.0500 M AgNO3 for 25.0 mL sample. Find [Cl-].
Working: 1:1 reaction. Moles Ag+ = 0.0500 x 0.0244 = 0.00122 mol = moles Cl-. [Cl-] = 0.00122/0.0250 = 0.0488 M.
Safety Precautions & Hazard Control
- AgNO3 stains; wear gloves and protect from light.
- Clean the Ag electrode before use.
- Switch off the power supply before changing any connections.
Waste Disposal Protocol
Collect AgCl precipitate and solutions for silver recovery.
Viva Voce Oral Defense Questions & Answers
Q: Why is there a sudden potential jump at the endpoint?
A: [Ag+] is pinned at ~10^-5 M by Ksp before the endpoint but rises freely with excess Ag+ after it - a huge logarithmic change.
Q: Why is no indicator needed?
A: The silver electrode itself signals the endpoint through the potential jump.