Potentiometric Acid-Base Titration with a pH Electrode
To find the endpoint of an HCl-NaOH titration from pH vs volume data.
Theoretical Foundations
pH changes slowly except near the endpoint, where it jumps ~6 units per drop. Plotting pH vs V gives a sigmoid; the endpoint is the inflection point, located precisely from the maximum of Delta(pH)/DeltaV.
Required Apparatus & Chemicals
Laboratory Apparatus
- pH meter with glass electrode
- Burette
- Beaker
- Magnetic stirrer
- Buffer solutions (pH 4 and 7)
Chemical Reagents
- Hydrochloric acid (~0.1 M)
- Sodium hydroxide (0.10 M standard)
- Distilled water (for rinsing glassware and electrodes)
Step-by-Step Practical Procedure
- Calibrate the pH meter with pH 7 and pH 4 buffers.
- Pipette 25.0 mL HCl into a beaker.
- Add NaOH in 1.0 mL steps (0.1 mL near the jump), recording pH.
- Plot pH vs V and Delta(pH)/DeltaV vs V.
- Read the endpoint from the derivative maximum and calculate [HCl].
Observations & Expected Results
pH jumps from ~4 to ~10 within 0.2 mL at the endpoint; derivative peak is sharp.
Worked Stoichiometric Calculations
Problem: Endpoint at 24.8 mL of 0.102 M NaOH for 25.0 mL HCl. Find [HCl].
Working: Moles NaOH = 0.102 x 0.0248 = 0.002530 mol. [HCl] = 0.002530/0.0250 = 0.101 M.
Safety Precautions & Hazard Control
- Handle the glass electrode bulb gently; keep it moist.
- Rinse the electrode between buffer and sample.
- Wear safety goggles and a lab coat throughout the experiment.
Waste Disposal Protocol
Neutralised solution can go down the drain.
Viva Voce Oral Defense Questions & Answers
Q: Why is the derivative plot better than the direct curve?
A: The inflection point is hard to read by eye on the sigmoid, but the Delta(pH)/DeltaV maximum locates it objectively.
Q: Why must the electrode be calibrated?
A: The glass electrode's asymmetry potential drifts; buffers fix the slope and offset.