DNA Denaturation: Hyperchromicity on Heating

To demonstrate DNA melting by the rise in 260 nm absorbance (hyperchromicity) on heating.

Level: UniversitySafety: SafeTime: 60 min
Aim of Experiment: To demonstrate DNA melting by the rise in 260 nm absorbance (hyperchromicity) on heating.

Theoretical Foundations

Stacked bases in duplex DNA absorb less at 260 nm; heating breaks hydrogen bonds and unstacks bases, raising absorbance ~30-40%. The midpoint is the melting temperature Tm, which rises with G-C content.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV spectrophotometer with heated cell holder or water bath
  • Quartz cuvettes
  • Cuvette tongs
  • Thermometer
  • Stopwatch
  • Ice bath

Chemical Reagents

  • DNA solution (A260 ~0.5)
  • TE buffer
  • Ice bath

Step-by-Step Practical Procedure

  1. Place DNA solution in a stoppered quartz cuvette and record A260 at 25 C.
  2. Warm stepwise (40, 60, 80, 95 C), equilibrating 5 minutes at each step, and record A260.
  3. Cool back to 25 C and record A260 again.
  4. Plot A260 versus temperature to see the melting transition.
  5. Calculate % hyperchromicity = (A95 - A25)/A25 x 100.

Observations & Expected Results

A260 stays flat, then climbs steeply between ~70-90 C; on cooling it remains high (strands do not readily reanneal).

Worked Stoichiometric Calculations

Problem: A260 = 0.50 at 25 C and 0.68 at 95 C. Calculate % hyperchromicity and estimate the transition midpoint.

Working: Hyperchromicity = (0.68 - 0.50)/0.50 x 100 = 36%. The midpoint (Tm) lies where A260 ~0.59, in the steep region of the plotted curve.

Safety Precautions & Hazard Control

  • Stopper cuvettes to prevent evaporation at high temperature.
  • Handle hot cuvettes with tissue/tweezers.
  • Equilibrate fully at each temperature.
  • Use quartz cuvettes only.

Waste Disposal Protocol

DNA-buffer waste can go down the drain.

Viva Voce Oral Defense Questions & Answers

Q: Why does absorbance rise when DNA melts?

A: Unstacking of bases removes hypochromic quenching, so each base absorbs more strongly at 260 nm.

Q: Why does rapid cooling not restore the original absorbance?

A: Separated strands reanneal slowly and imperfectly without controlled slow cooling; they stay largely single-stranded.