CHEMVEN

Protein Estimation by the Lowry Method

To estimate protein by the sensitive Lowry method combining biuret and Folin-Ciocalteu reactions.

Level: UniversitySafety: Needs supervisionTime: 90 min
Aim of Experiment: To estimate protein by the sensitive Lowry method combining biuret and Folin-Ciocalteu reactions.

Theoretical Foundations

Cu(II)-peptide complexes reduce the Folin-Ciocalteu phosphomolybdotungstate reagent to a blue complex (750 nm), amplified by tyrosine and tryptophan residues. It is ~100x more sensitive than biuret but slower.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Spectrophotometer
  • Test tubes
  • Pipettes
  • Stopwatch
  • Vortex mixer

Chemical Reagents

  • Lowry reagent (alkaline copper)
  • Folin-Ciocalteu reagent (1 N)
  • BSA standards (0-200 ug/mL)
  • Unknown sample

Step-by-Step Practical Procedure

  1. Add 5 mL Lowry reagent to 1 mL of each standard and diluted sample; stand 10 minutes.
  2. Add 0.5 mL Folin reagent rapidly with immediate mixing; start timing.
  3. Stand 30 minutes in the dark for full colour development.
  4. Read absorbance at 750 nm against the blank.
  5. Determine the sample protein from the standard curve.

Observations & Expected Results

A blue colour develops, deepest in concentrated standards; the curve is linear at low concentrations and bends at high ones.

Worked Stoichiometric Calculations

Problem: Linear region slope = 0.0045 A per ug/mL. A 1:50 diluted sample reads 0.54. Find the original protein concentration in mg/mL.

Working: Diluted C = 0.54/0.0045 = 120 ug/mL. Original = 120 x 50 = 6000 ug/mL = 6.0 mg/mL.

Safety Precautions & Hazard Control

  • Add Folin reagent quickly with instant mixing; timing affects colour.
  • Keep the 30-minute development identical for all tubes.
  • Folin reagent is acidic and corrosive; wear gloves.
  • Work in the linear range; dilute concentrated samples.

Waste Disposal Protocol

Folin waste (molybdenum/tungsten) collected as chemical waste.

Viva Voce Oral Defense Questions & Answers

Q: Why is Lowry more sensitive than biuret?

A: The Folin reduction by the copper-peptide complex plus aromatic residues greatly amplifies the colour per peptide bond.

Q: Name two common interferents in Lowry's method.

A: Reducing agents (e.g. DTT, mercaptoethanol) and detergents, chelators like EDTA, and high ammonium sulfate.

← Back to Experiments Directory