Folin-Wu Method: Blood Glucose Estimation (Demonstration)
To understand the classical Folin-Wu estimation of blood glucose using a simulated sample.
Theoretical Foundations
Proteins are precipitated with tungstic acid; the protein-free filtrate's glucose reduces alkaline copper tartrate, and the Cu2O formed reduces phosphomolybdic acid to molybdenum blue (read at 420 nm). Modern labs use enzymatic methods, but Folin-Wu teaches the redox principle.
Required Apparatus & Chemicals
Laboratory Apparatus
- Spectrophotometer
- Centrifuge
- Boiling water bath
- Pipettes
- Volumetric flasks
- Filter paper
Chemical Reagents
- Sodium tungstate (10%)
- Sulfuric acid (0.33 N)
- Alkaline copper tartrate
- Phosphomolybdic acid reagent
- Glucose standards
- Simulated serum sample
Step-by-Step Practical Procedure
- Mix 1 mL simulated sample with 7 mL water, 1 mL sodium tungstate and 1 mL dilute H2SO4 to precipitate proteins.
- Filter or centrifuge to obtain a clear protein-free filtrate.
- Heat 2 mL filtrate with 2 mL alkaline copper tartrate in a boiling bath for 8 minutes.
- Add 2 mL phosphomolybdic acid reagent; molybdenum blue develops.
- Dilute to 25 mL, read at 420 nm, and compare with glucose standards treated identically.
Observations & Expected Results
Standards and sample develop a blue colour proportional to glucose; the tungstic acid filtrate is crystal clear before the colour step.
Worked Stoichiometric Calculations
Problem: A standard containing 0.20 mg glucose in 2 mL gives A420 = 0.50. A 2 mL aliquot of 1:10 tungstic filtrate from simulated blood gives A420 = 0.45. What is the simulated blood glucose in mg/dL?
Working: Glucose in aliquot = 0.45/0.50 x 0.20 = 0.18 mg per 2 mL filtrate = 0.09 mg/mL filtrate. The filtrate is a 1:10 dilution of blood, so blood = 0.9 mg/mL = 90 mg/dL, a normal fasting value.
Safety Precautions & Hazard Control
- Use simulated samples, not real blood, in a teaching lab.
- Ensure complete protein precipitation for a clear filtrate.
- Time the 8-minute heating precisely.
- Phosphomolybdic acid is corrosive; wear gloves.
Waste Disposal Protocol
Neutralise acidic molybdate waste and collect molybdenum-containing liquid as chemical waste.
Viva Voce Oral Defense Questions & Answers
Q: Why must proteins be removed before the test?
A: Proteins would also reduce copper and give turbidity, falsely elevating and scattering the colour measurement.
Q: Why have enzymatic (GOD-POD) methods replaced Folin-Wu?
A: Glucose oxidase is specific for glucose, whereas copper reduction measures all reducing substances including other sugars and creatinine.