Biochemical Oxygen Demand (BOD) of Wastewater
To determine the 5-day BOD of a wastewater sample.
Theoretical Foundations
BOD5 is the oxygen consumed by microorganisms oxidising organic matter in 5 days at 20 C; it measures biodegradable pollution. Dissolved oxygen is fixed by the Winkler method (MnSO4 + alkaline KI, then acid) initially and after incubation, and titrated with thiosulphate. The difference, times the dilution factor, gives BOD in mg/L.
Required Apparatus & Chemicals
Laboratory Apparatus
- BOD bottles (300 mL)
- Incubator (20 C)
- Burette
- Pipettes
Chemical Reagents
- Wastewater sample (dilutions 1%, 5%, 25%)
- Dilution water (aerated, seeded)
- Manganous sulphate solution
- Alkaline potassium iodide
- Concentrated sulphuric acid
- 0.025 N sodium thiosulphate
- Starch indicator
Step-by-Step Practical Procedure
- Prepare dilutions of the sample in aerated dilution water (e.g. 1%, 5%, 25%).
- Fill BOD bottles completely, stopper without air bubbles.
- Fix DO in two bottles immediately (Winkler): add MnSO4, then alkaline KI, shake, settle, acidify.
- Titrate the liberated iodine with 0.025 N thiosulphate (DO initial).
- Incubate the remaining bottles 5 days at 20 C in the dark, then fix and titrate (DO final).
- Calculate BOD = (DO initial - DO final) x dilution factor.
Observations & Expected Results
Domestic sewage gives BOD 200-300 mg/L; treated effluent must be under 30 mg/L.
Worked Stoichiometric Calculations
Problem: A 5% dilution showed initial DO 8.0 mg/L and final DO 3.0 mg/L. Calculate the BOD5.
Working: Depletion = 8.0 - 3.0 = 5.0 mg/L. Dilution factor = 100/5 = 20. BOD5 = 5.0 x 20 = 100 mg/L.
Safety Precautions & Hazard Control
- Exclude all air bubbles when stoppering BOD bottles.
- Incubate in complete darkness to stop photosynthesis.
- Use seeded dilution water for industrial wastes lacking microbes.
- Keep incubation at exactly 20 C.
Waste Disposal Protocol
Titrated samples go down the drain; incubated bottles are disinfected before disposal.
Viva Voce Oral Defense Questions & Answers
Q: Why 5 days at 20 C?
A: The standard adopted historically: 5 days was the longest river travel time in England, and 20 C is a typical summer river temperature.
Q: What is the Winkler method?
A: Dissolved oxygen oxidises Mn(OH)2 to MnO2, which in acid liberates iodine from KI, titrated with thiosulphate.