Carbon Dioxide in Ambient and Indoor Air: NDIR Measurement

To measure CO2 in outdoor and indoor air with a non-dispersive infrared (NDIR) sensor.

Level: UniversitySafety: SafeTime: 30 min
Aim of Experiment: To measure CO2 in outdoor and indoor air with a non-dispersive infrared (NDIR) sensor.

Theoretical Foundations

NDIR sensors measure CO2 absorption of infrared at 4.26 um; absorbance is proportional to concentration. Outdoor background is ~420 ppm (2024); indoor levels above 1000 ppm indicate poor ventilation. CO2 itself is a ventilation indicator indoors and the principal greenhouse gas outdoors.

Required Apparatus & Chemicals

Laboratory Apparatus

  • NDIR CO2 meter (calibrated)
  • Notebook for site conditions
  • Stopwatch
  • Tripod/stand for the sensor

Chemical Reagents

  • No chemicals (direct sensor reading)
  • Calibration gas (span check, if available)
  • Silica gel (sensor storage)

Step-by-Step Practical Procedure

  1. Calibrate/zero the meter in fresh outdoor air if required.
  2. Record outdoor background CO2.
  3. Measure in a classroom/lecture hall before occupancy.
  4. Measure again at peak occupancy.
  5. Measure near a busy road.
  6. Compare with the 1000 ppm indoor ventilation guideline and note the traffic increment.

Observations & Expected Results

Outdoor: ~420 ppm; crowded poorly ventilated rooms: 1500-2500 ppm; roadside: 450-600 ppm.

Worked Stoichiometric Calculations

Problem: A classroom reads 1800 ppm CO2 at peak occupancy against 420 ppm outdoors. Calculate the indoor increment and comment.

Working: Increment = 1800 - 420 = 1380 ppm. Well above the ~600 ppm above-background ventilation guideline (1000 ppm absolute); ventilation must be improved.

Safety Precautions & Hazard Control

  • Allow the sensor to stabilise at each site.
  • Avoid breathing directly onto the sensor.
  • Record occupancy and ventilation status.
  • Calibrate periodically against fresh air.

Waste Disposal Protocol

No waste generated.

Viva Voce Oral Defense Questions & Answers

Q: Why is indoor CO2 a ventilation indicator?

A: Humans exhale CO2 at a steady rate, so its build-up directly reflects inadequate fresh-air exchange.

Q: What is the current outdoor background CO2?

A: About 420 ppm and rising ~2.5 ppm per year from fossil-fuel combustion.

Q: At what levels does CO2 itself impair cognition?

A: Studies show measurable cognitive decline above ~1000-1400 ppm, with drowsiness and headaches higher still.

← Back to Experiments Directory