PM2.5 and PM10: Size-Selective Ambient Sampling
To determine PM10 and PM2.5 concentrations with size-selective inlets and compare with air quality standards.
Theoretical Foundations
Cyclone/impactor inlets separate particles by aerodynamic diameter before collection on filters: PM10 inlet passes particles <10 um, PM2.5 inlet <2.5 um. Gravimetric analysis as for SPM gives each fraction. Coarse fraction (PM10-PM2.5) indicates dust/mechanical sources; the fine fraction indicates combustion sources.
Required Apparatus & Chemicals
Laboratory Apparatus
- PM10 and PM2.5 samplers (or one sampler with interchangeable inlets)
- Pre-weighed PTFE/quartz filters
- Microbalance
- Desiccator
Chemical Reagents
- Silica gel (desiccant)
- Distilled water (inlet cleaning)
- Isopropanol (grease removal)
Step-by-Step Practical Procedure
- Condition and weigh filters for both inlets.
- Deploy PM10 and PM2.5 samplers side by side for 24 hours.
- Record flow rates and sampling duration.
- Condition and reweigh filters.
- Compute PM10 and PM2.5 in ug/m3.
- Calculate the coarse fraction (PM10 - PM2.5) and PM2.5/PM10 ratio; compare with standards.
Observations & Expected Results
Traffic sites: PM2.5/PM10 ratio 0.5-0.7; desert-dust episodes: ratio <0.3 with high coarse fraction.
Worked Stoichiometric Calculations
Problem: PM10 = 132 ug/m3, PM2.5 = 78 ug/m3. Find the coarse fraction and ratio, and check NAAQS (PM10 100, PM2.5 60 ug/m3).
Working: Coarse = 132 - 78 = 54 ug/m3. Ratio PM2.5/PM10 = 78/132 = 0.59, indicating significant combustion sources. Both exceed NAAQS (100 and 60 ug/m3); the air quality is poor.
Safety Precautions & Hazard Control
- Use the correct inlet for each fraction; do not interchange.
- Maintain the design flow rate; cut-point depends on it.
- Avoid filter overloading.
- Keep inlets clean.
Waste Disposal Protocol
Used filters to solid waste.
Viva Voce Oral Defense Questions & Answers
Q: What does the PM2.5/PM10 ratio tell us?
A: A high ratio (>0.5) points to combustion sources; a low ratio points to mechanical dust sources.
Q: What is aerodynamic diameter?
A: The diameter of a unit-density sphere with the same settling velocity as the particle; it governs sampler cut-points.
Q: Why does the cut-point depend on flow rate?
A: Cyclone/impactor separation is inertial; changing flow changes particle inertia and hence the size collected.