Dissolved Organic Carbon Proxy: UV254 Absorbance (SUVA)
To estimate dissolved organic carbon (DOC) in water using UV absorbance at 254 nm (SUVA concept).
Theoretical Foundations
Aromatic organic matter absorbs UV at 254 nm; UV254 absorbance correlates with DOC, and SUVA (UV254/DOC) indicates aromaticity and disinfection-by-product formation potential. Filtered samples are measured in quartz cuvettes. High-SUVA waters form more trihalomethanes on chlorination.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV spectrophotometer
- Quartz cuvettes
- Membrane filter 0.45 um
- Filtration assembly
- Wash bottle
Chemical Reagents
- Potassium hydrogen phthalate (DOC standards, optional)
- Distilled water
- Dilute sulphuric acid (pH adjustment, if needed)
Step-by-Step Practical Procedure
- Filter the sample through a 0.45 um membrane.
- Zero the spectrophotometer with distilled water at 254 nm.
- Measure sample absorbance at 254 nm in a quartz cuvette.
- If DOC is known (TOC analyser), compute SUVA = UV254 x 100 / DOC.
- Use the site-specific UV254-DOC correlation to estimate DOC.
- Assess disinfection-by-product formation potential.
Observations & Expected Results
Humic-rich waters: UV254 0.1-0.5 cm-1; clear groundwater: <0.02 cm-1.
Worked Stoichiometric Calculations
Problem: A filtered sample has UV254 = 0.12 cm-1 and DOC = 4.0 mg/L. Calculate SUVA and interpret.
Working: SUVA = 0.12 x 100 / 4.0 = 3.0 L/mg-m. SUVA >2 indicates aromatic, humic-rich organics with significant trihalomethane formation potential on chlorination; enhanced coagulation is advisable.
Safety Precautions & Hazard Control
- Filter before measurement; particles scatter UV.
- Use quartz cuvettes.
- Correct for nitrate interference if high (absorbs near 220 nm, minimal at 254).
- Keep pH consistent; UV absorbance is pH-sensitive.
Waste Disposal Protocol
No hazardous waste; aqueous samples only.
Viva Voce Oral Defense Questions & Answers
Q: What is SUVA?
A: Specific UV absorbance = UV254 (per metre) divided by DOC (mg/L); a measure of aromatic carbon content.
Q: Why does high SUVA matter for chlorination?
A: Aromatic organics are the main precursors of trihalomethanes and haloacetic acids.
Q: Why filter at 0.45 um?
A: To measure dissolved organics only; particles would scatter light and inflate absorbance.