CHEMVEN

Arsenic in Groundwater: Silver Diethyldithiocarbamate Method

To determine trace arsenic in groundwater by arsine generation and SDDC colorimetry.

Level: UniversitySafety: Advanced lab onlyTime: 2 h
Aim of Experiment: To determine trace arsenic in groundwater by arsine generation and SDDC colorimetry.

Theoretical Foundations

Arsenic is reduced to arsine gas (AsH3) by zinc in acid; the gas is scrubbed of H2S by lead acetate cotton and absorbed in silver diethyldithiocarbamate (SDDC) in pyridine, forming a red complex measured at 535 nm. WHO guideline: 0.01 mg/L; chronic exposure causes skin lesions and cancers. The method detects down to ~0.002 mg/L.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Arsine generator (Gutzeit-type bottle with scrubber tube)
  • Spectrophotometer (535 nm)
  • Cuvettes
  • Fume hood

Chemical Reagents

  • Zinc granules (arsenic-free)
  • Hydrochloric acid (conc.)
  • Lead acetate impregnated cotton
  • SDDC reagent in pyridine
  • Sodium arsenite stock standard
  • Potassium iodide + stannous chloride (reductants)

Step-by-Step Practical Procedure

  1. Take 50 mL of acid-preserved sample in the generator bottle.
  2. Add KI and stannous chloride to reduce As(V) to As(III); wait 15 minutes.
  3. Place lead acetate cotton in the scrubber tube and SDDC-pyridine in the absorber.
  4. Add zinc granules and conc. HCl; connect the apparatus immediately.
  5. Allow 45 minutes for complete arsine evolution and absorption (red colour develops).
  6. Measure absorbance at 535 nm against a reagent blank.
  7. Read arsenic from a calibration curve run through the same generation step.

Observations & Expected Results

Yellow to red colour in the absorber. Arsenic-affected aquifers (Bengal delta type): 0.05-1.0 mg/L, far above the 0.01 mg/L WHO guideline.

Worked Stoichiometric Calculations

Problem: A 50 mL sample gives A = 0.30 at 535 nm; calibration A = 6.0 x C (mg/L As). Find arsenic and comment.

Working: C = 0.30/6.0 = 0.05 mg/L As. This is five times the WHO guideline of 0.01 mg/L; the well must not be used for drinking without arsenic removal treatment.

Safety Precautions & Hazard Control

  • Arsine is extremely toxic; assemble quickly and work in a fume hood.
  • The lead acetate scrubber is essential to remove H2S, which also colours SDDC.
  • Use arsenic-free zinc and acids; run blanks.
  • Pyridine is toxic and foul-smelling; handle in the fume hood.

Waste Disposal Protocol

SDDC-pyridine absorber solutions and generator residues go to the hazardous waste container.

Viva Voce Oral Defense Questions & Answers

Q: What is the function of the lead acetate cotton?

A: It scrubs hydrogen sulphide from the gas stream; H2S would also react with SDDC and give false high arsenic.

Q: Why are KI and stannous chloride added first?

A: They pre-reduce As(V) to As(III), which is the form that generates arsine with zinc and acid.

Q: Name a chronic health effect of arsenic in drinking water.

A: Hyperkeratosis/skin lesions, and long-term cancers of skin, bladder and lung.

← Back to Experiments Directory