CHEMVEN

Sulphide in Water: Iodometric Titration

To determine dissolved sulphide in wastewater by iodometric titration.

Level: UniversitySafety: Needs supervisionTime: 60 min
Aim of Experiment: To determine dissolved sulphide in wastewater by iodometric titration.

Theoretical Foundations

Sulphide is fixed at sampling with zinc acetate as ZnS. After acidification, the sulphide reduces a known excess of iodine, and the remaining iodine is back-titrated with thiosulphate. Sulphide (mg/L) = (B - S) x N x 16000 / V. Sulphide is acutely toxic to fish above ~0.05 mg/L and generates H2S odour and concrete corrosion in sewers.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Iodine flasks
  • Burette
  • Pipettes
  • Measuring cylinder

Chemical Reagents

  • Zinc acetate solution (fixative)
  • Iodine solution (0.025 N)
  • Sodium thiosulphate (0.025 N)
  • Hydrochloric acid (1:1)
  • Starch indicator

Step-by-Step Practical Procedure

  1. Fix the sample at collection by adding zinc acetate (white ZnS precipitates).
  2. Take a measured volume, add a known excess of 0.025 N iodine solution.
  3. Acidify with HCl; the ZnS dissolves and reduces iodine.
  4. Back-titrate residual iodine with 0.025 N thiosulphate using starch.
  5. Run a blank with iodine alone.
  6. Calculate sulphide = (B - S) x N x 16000 / V mg/L.

Observations & Expected Results

Rotten-egg odour; white ZnS precipitate. Tannery/sewer samples: 1-50 mg/L sulphide.

Worked Stoichiometric Calculations

Problem: 100 mL of fixed sample: blank titre 20.0 mL, sample titre 14.5 mL of 0.025 N thiosulphate. Calculate sulphide in mg/L.

Working: Sulphide = (20.0 - 14.5) x 0.025 x 16000 / 100 = 5.5 x 400 / 100 = 22 mg/L. Highly toxic to aquatic life; the effluent needs oxidation treatment (e.g. aeration/chlorination) before discharge.

Safety Precautions & Hazard Control

  • Fix sulphide immediately at sampling; H2S escapes rapidly.
  • Work in a ventilated area; H2S is toxic.
  • Acidify cautiously to avoid H2S release into the lab.
  • Standardise iodine and thiosulphate on the same day.

Waste Disposal Protocol

Zinc-containing acidic waste: neutralise and collect in the inorganic waste container.

Viva Voce Oral Defense Questions & Answers

Q: Why is zinc acetate added at sampling?

A: It precipitates sulphide as stable ZnS, preventing loss of volatile H2S before analysis.

Q: Derive the factor 16000.

A: Equivalent weight of S2- = 32.06/2 = 16.03 g/equiv; mg/L = V x N x 16.03 x 1000 / V_sample, i.e. V x N x 16030 / V, conventionally 16000.

Q: Why is sulphide a sewer problem?

A: Bacteria oxidise H2S to sulphuric acid on sewer crowns, corroding concrete (crown corrosion).

← Back to Experiments Directory