Oil and Grease in Wastewater: Hexane Extraction Gravimetry

To determine oil and grease in wastewater by n-hexane extraction and gravimetry.

Level: UniversitySafety: Advanced lab onlyTime: 2 h
Aim of Experiment: To determine oil and grease in wastewater by n-hexane extraction and gravimetry.

Theoretical Foundations

Oil and grease (hydrocarbons, fats, waxes) are extracted from an acidified sample with n-hexane in a separating funnel; the solvent is evaporated and the residue weighed. Acidification to pH <2 hydrolyses soaps and releases emulsified oil. Discharge standards commonly limit oil and grease to 10 mg/L; they form surface films that block oxygen transfer in rivers.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Separating funnel 1 L
  • Evaporating dish (weighed)
  • Water bath
  • Desiccator
  • Analytical balance
  • Fume hood

Chemical Reagents

  • n-Hexane (analytical grade)
  • Hydrochloric acid (1:1)
  • Anhydrous sodium sulphate

Step-by-Step Practical Procedure

  1. Acidify 1 L of sample to pH <2 with HCl in a separating funnel.
  2. Add 30 mL n-hexane, shake vigorously for 2 minutes, and let the layers separate.
  3. Drain the aqueous layer; filter the hexane layer through anhydrous sodium sulphate into a weighed dish.
  4. Repeat extraction twice more with fresh hexane.
  5. Evaporate the hexane on a water bath in a fume hood.
  6. Dry the dish at 70 C, cool in a desiccator, and weigh to constant weight.
  7. Oil and grease (mg/L) = (dish gain in mg) / sample volume in L.

Observations & Expected Results

Oily residue with characteristic odour. Canteen/automobile effluents: 50-500 mg/L; discharge limit commonly 10 mg/L.

Worked Stoichiometric Calculations

Problem: 1 L of effluent gave a dish weight gain of 86 mg. Calculate oil and grease and check a 10 mg/L discharge limit.

Working: Oil and grease = 86 mg / 1 L = 86 mg/L. This exceeds the 10 mg/L discharge limit nearly nine-fold; an oil separator or dissolved-air flotation unit is required.

Safety Precautions & Hazard Control

  • Work entirely in a fume hood; hexane is highly flammable and its vapours narcotic.
  • No open flames anywhere near the extraction.
  • Dry to constant weight; residual solvent inflates results.
  • Acidify promptly at sampling to stop biodegradation.

Waste Disposal Protocol

Recover distilled hexane for reuse; oily residues go to the hazardous organic waste container.

Viva Voce Oral Defense Questions & Answers

Q: Why is the sample acidified before extraction?

A: Acid hydrolyses soaps and breaks oil emulsions, releasing bound oil into the solvent.

Q: Why is sodium sulphate used in filtration?

A: Anhydrous sodium sulphate removes traces of water from the hexane extract so only oil and grease are weighed.

Q: Why is oil and grease an environmental concern?

A: It forms surface films that block atmospheric oxygen transfer and coats aquatic organisms and treatment-plant biomass.