Iron in Water: 1,10-Phenanthroline Method
To determine dissolved iron in water by the 1,10-phenanthroline colorimetric method.
Theoretical Foundations
Hydroxylamine reduces Fe3+ to Fe2+, which forms an orange-red complex with 1,10-phenanthroline at pH 3-4 (acetate buffer), measured at 510 nm. WHO guideline: 0.3 mg/L (taste and staining, not health); iron causes laundry staining and metallic taste and promotes iron-bacteria growth in pipes.
Required Apparatus & Chemicals
Laboratory Apparatus
- Spectrophotometer (510 nm)
- Cuvettes
- Volumetric flasks 50 mL
- Pipettes
- Hot plate (for digestion, if needed)
Chemical Reagents
- 1,10-phenanthroline solution
- Hydroxylamine hydrochloride solution
- Sodium acetate buffer
- Ferrous ammonium sulphate stock (Fe standard)
- Hydrochloric acid (for preservation)
Step-by-Step Practical Procedure
- Preserve the sample with HCl to pH <2 at collection.
- Prepare iron standards 0-1.0 mg/L Fe.
- To 50 mL of each standard and sample add hydroxylamine, acetate buffer and phenanthroline.
- Wait 10 minutes for full colour development.
- Measure absorbance at 510 nm against a reagent blank.
- Read iron from the calibration curve and compare with the 0.3 mg/L WHO guideline.
Observations & Expected Results
Orange-red colour proportional to iron. Ferruginous groundwater: 1-10 mg/L; aeration precipitates brown Fe(OH)3.
Worked Stoichiometric Calculations
Problem: A tube-well sample gives A = 0.36 at 510 nm. Calibration: A = 0.90 x C (mg/L Fe). Find iron and comment.
Working: C = 0.36/0.90 = 0.40 mg/L Fe. This exceeds the WHO guideline of 0.3 mg/L; aeration-filtration treatment is recommended to avoid staining and taste.
Safety Precautions & Hazard Control
- Preserve samples with acid immediately; iron oxidises and precipitates on standing.
- Keep pH 3-4 during colour development.
- Avoid phosphate and strong oxidants that interfere.
- Use iron-free distilled water for the blank.
Waste Disposal Protocol
Phenanthroline waste goes to the organic chemical waste container.
Viva Voce Oral Defense Questions & Answers
Q: Why is hydroxylamine added?
A: It reduces all Fe3+ to Fe2+, because phenanthroline complexes only the ferrous form.
Q: Why must iron samples be acidified at collection?
A: Acid keeps iron dissolved; at neutral pH Fe2+ oxidises to Fe3+ and precipitates as Fe(OH)3, giving low results.
Q: Is the WHO iron limit health-based?
A: No; 0.3 mg/L is based on taste and staining of laundry and plumbing.