Phosphate in Water: Molybdenum Blue Method
To determine orthophosphate in water by the ascorbic acid-molybdenum blue method.
Theoretical Foundations
Orthophosphate reacts with ammonium molybdate in acid to form phosphomolybdic acid, which ascorbic acid reduces to intensely blue molybdenum blue, measured at 880 nm. It is the key nutrient assay for eutrophication studies; phosphate above ~0.1 mg/L as P can trigger algal blooms in lakes. Total phosphorus needs prior persulphate digestion of organic and condensed phosphates.
Required Apparatus & Chemicals
Laboratory Apparatus
- Spectrophotometer (880 nm)
- Cuvettes
- Volumetric flasks
- Water bath (for digestion, optional)
Chemical Reagents
- Ammonium molybdate solution
- Sulphuric acid (5 N)
- Ascorbic acid solution (fresh)
- Potassium dihydrogen phosphate stock
- Potassium persulphate (for total P digestion)
Step-by-Step Practical Procedure
- Prepare phosphate standards 0-1.0 mg/L as P.
- Add combined molybdate-acid reagent and ascorbic acid to standards and samples.
- Allow 10 minutes for full blue colour development.
- Measure absorbance at 880 nm against a reagent blank.
- Plot the calibration curve and read sample phosphate.
- For total phosphorus, digest a separate portion with persulphate first, then develop colour.
Observations & Expected Results
Colourless to deep blue proportional to phosphate. Eutrophic lake water: >0.1 mg/L as P; pristine streams: <0.02 mg/L as P.
Worked Stoichiometric Calculations
Problem: A lake sample gives A = 0.55 at 880 nm. Calibration: A = 1.10 x C (C in mg/L as P). Find phosphate and assess eutrophication risk.
Working: C = 0.55/1.10 = 0.50 mg/L as P. This is five times the ~0.1 mg/L bloom threshold, indicating a high eutrophication risk.
Safety Precautions & Hazard Control
- Prepare ascorbic acid fresh daily; it oxidises in air.
- Keep the development time constant for all flasks.
- Silica above 50 mg/L gives a similar blue colour; keep pH controlled.
- Acid-wash glassware to remove adsorbed phosphate.
Waste Disposal Protocol
Molybdenum-containing waste goes to the inorganic waste container.
Viva Voce Oral Defense Questions & Answers
Q: Why is 880 nm used instead of visible wavelengths?
A: Molybdenum blue absorbs maximally near 880 nm, giving the highest sensitivity and least interference.
Q: How does total phosphorus differ from orthophosphate?
A: Total P includes organic and condensed phosphates, released by persulphate digestion; orthophosphate is the immediately bioavailable dissolved fraction.
Q: Why is phosphate the usual limiting nutrient in fresh waters?
A: Nitrogen is often abundant relative to demand, so algal growth is typically capped by available phosphorus.