Turbidity of Water: Nephelometric Measurement
To determine the turbidity of water samples using a nephelometer in NTU.
Theoretical Foundations
Turbidity is the cloudiness caused by suspended particles, measured by the light scattered at 90 degrees to an incident beam (nephelometry) in Nephelometric Turbidity Units (NTU). The instrument is calibrated with formazin standards. WHO guideline: 1 NTU for effective disinfection (5 NTU maximum); turbidity shields pathogens from chlorine.
Required Apparatus & Chemicals
Laboratory Apparatus
- Nephelometer/turbidity meter
- Sample cuvettes
- Beakers
- Wash bottle
- Lint-free wipes
Chemical Reagents
- Formazin turbidity standards (0, 10, 100 NTU)
- Distilled (turbidity-free) water
- Cuvette cleaning solution
Step-by-Step Practical Procedure
- Calibrate the nephelometer with 0, 10 and 100 NTU formazin standards.
- Shake the sample gently to resuspend particles without creating bubbles.
- Fill a clean cuvette, wipe its outside dry, and insert it into the meter.
- Record the NTU reading once stable.
- Repeat for each sample; rinse cuvettes between readings.
- Compare results with the WHO guideline of 1 NTU.
Observations & Expected Results
Clear tap water: <1 NTU; river water in monsoon: 50-500 NTU; settled/filtered water: <1 NTU.
Worked Stoichiometric Calculations
Problem: A filtered water sample reads 0.8 NTU and the raw river water reads 120 NTU. Calculate the percentage turbidity removal and check WHO compliance.
Working: Removal = (120 - 0.8)/120 x 100 = 99.33%. Filtered water at 0.8 NTU meets the WHO guideline of 1 NTU.
Safety Precautions & Hazard Control
- Avoid air bubbles in the cuvette; they scatter light and inflate readings.
- Clean cuvettes; scratches and fingerprints cause errors.
- Analyse promptly; particles settle on standing.
- Recalibrate if the instrument has been moved.
Waste Disposal Protocol
Formazin standards contain hydrazine sulphate; collect used standards in chemical waste, not the drain.
Viva Voce Oral Defense Questions & Answers
Q: Why is turbidity important for disinfection?
A: Suspended particles shield bacteria and viruses from chlorine/UV, so high turbidity makes disinfection ineffective.
Q: What is the principle of nephelometry?
A: Measuring light scattered at 90 degrees by suspended particles, proportional to particle concentration.
Q: Why must cuvettes be free of scratches?
A: Scratches scatter light like particles do, giving falsely high turbidity readings.