pH of Water Samples: Electrometric Determination
To measure the pH of various water samples with a calibrated pH meter and assess compliance with drinking-water standards.
Theoretical Foundations
pH is the negative logarithm of hydrogen ion activity and is measured with a glass electrode against a reference electrode. The meter is calibrated with standard buffers (pH 4.0, 7.0, 9.2) with temperature compensation. WHO drinking-water guideline range is pH 6.5-8.5; outside this range water may corrode pipes or indicate pollution.
Required Apparatus & Chemicals
Laboratory Apparatus
- pH meter with glass electrode
- Beakers 100 mL
- Thermometer
- Wash bottle
Chemical Reagents
- Buffer tablets/solutions pH 4.0, 7.0, 9.2
- Distilled water
- Potassium chloride (electrode storage solution)
Step-by-Step Practical Procedure
- Switch on the pH meter and allow it to warm up.
- Rinse the electrode with distilled water and blot dry.
- Calibrate with pH 7.0 buffer, then check with pH 4.0 (or 9.2) buffer; adjust slope if needed.
- Rinse the electrode and immerse it in the first water sample.
- Stir gently, wait for a stable reading, and record pH and temperature.
- Repeat for each sample, rinsing between measurements.
- Store the electrode in KCl solution after use.
Observations & Expected Results
Stable readings within 1-2 minutes. Typical: tap water 7.0-7.8, rainwater 5.6 (natural CO2), acidic industrial effluent below 6.5. WHO standard: 6.5-8.5.
Worked Stoichiometric Calculations
Problem: A sample has pH 6.2. Calculate its hydrogen ion concentration and state whether it meets the WHO drinking-water range.
Working: [H+] = 10^-6.2 = 6.31 x 10^-7 mol/L. It lies below the WHO lower limit of 6.5, so it does not meet the guideline (likely corrosive).
Safety Precautions & Hazard Control
- Calibrate daily and check slope with two buffers.
- Never let the glass bulb dry out; keep it in storage solution.
- Apply temperature compensation for accurate readings.
- Rinse between samples to avoid carryover.
Waste Disposal Protocol
Buffer solutions are harmless and may go down the drain.
Viva Voce Oral Defense Questions & Answers
Q: Define pH.
A: pH = -log10 a(H+), the negative logarithm of hydrogen ion activity.
Q: Why are two buffers used for calibration?
A: One sets the zero point (pH 7) and the second sets the electrode slope, correcting for non-ideal electrode response.
Q: Why does natural rainwater have pH about 5.6?
A: Dissolved atmospheric CO2 forms carbonic acid, giving unpolluted rain a slightly acidic pH near 5.6.