pH of Water Samples: Electrometric Determination

To measure the pH of various water samples with a calibrated pH meter and assess compliance with drinking-water standards.

Level: UniversitySafety: SafeTime: 30 min
Aim of Experiment: To measure the pH of various water samples with a calibrated pH meter and assess compliance with drinking-water standards.

Theoretical Foundations

pH is the negative logarithm of hydrogen ion activity and is measured with a glass electrode against a reference electrode. The meter is calibrated with standard buffers (pH 4.0, 7.0, 9.2) with temperature compensation. WHO drinking-water guideline range is pH 6.5-8.5; outside this range water may corrode pipes or indicate pollution.

Required Apparatus & Chemicals

Laboratory Apparatus

  • pH meter with glass electrode
  • Beakers 100 mL
  • Thermometer
  • Wash bottle

Chemical Reagents

  • Buffer tablets/solutions pH 4.0, 7.0, 9.2
  • Distilled water
  • Potassium chloride (electrode storage solution)

Step-by-Step Practical Procedure

  1. Switch on the pH meter and allow it to warm up.
  2. Rinse the electrode with distilled water and blot dry.
  3. Calibrate with pH 7.0 buffer, then check with pH 4.0 (or 9.2) buffer; adjust slope if needed.
  4. Rinse the electrode and immerse it in the first water sample.
  5. Stir gently, wait for a stable reading, and record pH and temperature.
  6. Repeat for each sample, rinsing between measurements.
  7. Store the electrode in KCl solution after use.

Observations & Expected Results

Stable readings within 1-2 minutes. Typical: tap water 7.0-7.8, rainwater 5.6 (natural CO2), acidic industrial effluent below 6.5. WHO standard: 6.5-8.5.

Worked Stoichiometric Calculations

Problem: A sample has pH 6.2. Calculate its hydrogen ion concentration and state whether it meets the WHO drinking-water range.

Working: [H+] = 10^-6.2 = 6.31 x 10^-7 mol/L. It lies below the WHO lower limit of 6.5, so it does not meet the guideline (likely corrosive).

Safety Precautions & Hazard Control

  • Calibrate daily and check slope with two buffers.
  • Never let the glass bulb dry out; keep it in storage solution.
  • Apply temperature compensation for accurate readings.
  • Rinse between samples to avoid carryover.

Waste Disposal Protocol

Buffer solutions are harmless and may go down the drain.

Viva Voce Oral Defense Questions & Answers

Q: Define pH.

A: pH = -log10 a(H+), the negative logarithm of hydrogen ion activity.

Q: Why are two buffers used for calibration?

A: One sets the zero point (pH 7) and the second sets the electrode slope, correcting for non-ideal electrode response.

Q: Why does natural rainwater have pH about 5.6?

A: Dissolved atmospheric CO2 forms carbonic acid, giving unpolluted rain a slightly acidic pH near 5.6.