Chemical Oxygen Demand: Dichromate Open-Reflux Method
To determine the chemical oxygen demand (COD) of a wastewater sample by potassium dichromate open reflux.
Theoretical Foundations
COD is the oxygen equivalent of the organic (and some inorganic) matter oxidisable by hot acidic potassium dichromate, the standard measure of total pollution load. The sample is refluxed for 2 hours with excess K2Cr2O7 in concentrated H2SO4 with silver sulphate catalyst; mercuric sulphate complexes chloride to prevent its oxidation. Unreacted dichromate is back-titrated with ferrous ammonium sulphate (FAS) using ferroin indicator. COD (mg/L) = (B - S) x N x 8000 / V, where B and S are blank and sample titres.
Required Apparatus & Chemicals
Laboratory Apparatus
- Reflux apparatus (round-bottom flask + condenser)
- Heating mantle
- Burette 50 mL
- Pipettes
- Anti-bumping granules
- Conical flasks
Chemical Reagents
- Potassium dichromate (0.25 N)
- Concentrated sulphuric acid
- Silver sulphate in conc. H2SO4 (catalyst)
- Mercuric sulphate (solid)
- Ferrous ammonium sulphate (0.1 N)
- Ferroin indicator
Step-by-Step Practical Procedure
- Place 20 mL of well-mixed sample (or diluted sample) in the reflux flask.
- Add 0.4 g mercuric sulphate and swirl to dissolve.
- Add 10 mL of 0.25 N potassium dichromate solution.
- Cautiously add 30 mL of conc. H2SO4 containing silver sulphate, cooling the flask.
- Add anti-bumping granules and reflux gently for 2 hours.
- Cool, wash down the condenser with 25 mL distilled water.
- Titrate the cooled digest with 0.1 N FAS using 2 drops ferroin to a sharp reddish-brown endpoint.
- Run a blank with distilled water through the same procedure and calculate COD.
Observations & Expected Results
Digest changes from orange (excess dichromate) toward green (Cr3+); ferroin endpoint is sharp blue-green to reddish brown. Raw sewage COD 250-800 mg/L; treated effluent discharge standard commonly 250 mg/L (industrial). COD is always greater than or equal to BOD for the same sample.
Worked Stoichiometric Calculations
Problem: 20 mL of effluent required 12.4 mL of 0.1 N FAS; the blank required 24.8 mL. Calculate COD in mg/L.
Working: COD = (B - S) x N x 8000 / V = (24.8 - 12.4) x 0.1 x 8000 / 20 = 12.4 x 800 / 20 = 12.4 x 40 = 496 mg/L.
Safety Precautions & Hazard Control
- Add concentrated acid very slowly with cooling; the mixture gets extremely hot.
- Mercuric sulphate and hexavalent chromium are toxic; wear gloves and avoid inhalation.
- Ensure steady, gentle reflux without violent bumping.
- Never pour dichromate or mercury waste down the drain.
Waste Disposal Protocol
Collect the chromium- and mercury-containing digest waste in the labelled heavy-metal waste container for authorised disposal.
Viva Voce Oral Defense Questions & Answers
Q: Why is silver sulphate added in the COD test?
A: Silver ion catalyses the oxidation of straight-chain aliphatic compounds, which dichromate alone oxidises poorly.
Q: Why is mercuric sulphate added?
A: It complexes chloride ions as soluble HgCl2, preventing chloride from being oxidised to chlorine and giving falsely high COD.
Q: Why is COD usually higher than BOD?
A: Dichromate oxidises nearly all organic matter including non-biodegradable fractions, while BOD measures only the biodegradable part.