Biochemical Oxygen Demand: 5-Day BOD Test

To determine the 5-day biochemical oxygen demand (BOD5) of a wastewater sample at 20 C.

Level: UniversitySafety: Needs supervisionTime: 2 h (plus 5-day incubation)
Aim of Experiment: To determine the 5-day biochemical oxygen demand (BOD5) of a wastewater sample at 20 C.

Theoretical Foundations

BOD is the amount of oxygen consumed by microorganisms while biodegrading organic matter in a sample over 5 days at 20 C in the dark (BOD5). It is the standard index of biodegradable pollution load. The sample is diluted with seeded, oxygen-saturated dilution water, and DO is measured before and after incubation by Winkler titration. BOD (mg/L) = (D1 - D2) x dilution factor. Domestic sewage typically shows BOD 100-300 mg/L; treated effluent should be below 30 mg/L (common discharge standard).

Required Apparatus & Chemicals

Laboratory Apparatus

  • BOD bottles 300 mL (several)
  • BOD incubator at 20 C
  • Burette, pipettes
  • Measuring cylinder
  • Air pump/aerator for dilution water
  • Thermometer

Chemical Reagents

  • Phosphate buffer solution
  • Magnesium sulphate solution
  • Calcium chloride solution
  • Ferric chloride solution
  • Seeded dilution water (distilled water + nutrients + settled sewage seed)
  • Winkler reagents (manganous sulphate, alkali-iodide-azide, conc. H2SO4)
  • Sodium thiosulphate (0.025 N)
  • Starch indicator

Step-by-Step Practical Procedure

  1. Prepare dilution water: aerate distilled water, add 1 mL each of phosphate buffer, MgSO4, CaCl2 and FeCl3 per litre, then add settled sewage seed.
  2. Prepare at least three dilutions of the wastewater (e.g. 1%, 2%, 5%) in BOD bottles filled to the brim.
  3. Fill two BOD bottles with dilution water alone as blanks.
  4. Determine initial DO (D1) of one bottle of each dilution and one blank by Winkler titration.
  5. Incubate the remaining bottles in the dark at 20 C for 5 days.
  6. After incubation, determine final DO (D2) of each bottle.
  7. Select a dilution where D1 - D2 is between 2 and 7 mg/L and at least 1 mg/L residual DO remains.
  8. Calculate BOD5 = (D1 - D2) x dilution factor, correcting for blank depletion if needed.

Observations & Expected Results

DO falls during incubation; healthy dilution water blanks deplete less than 0.2 mg/L. Untreated domestic sewage: BOD5 100-300 mg/L. Treated sewage discharge standard (common): 30 mg/L.

Worked Stoichiometric Calculations

Problem: A 2% dilution of wastewater had initial DO 8.6 mg/L and final DO 4.1 mg/L after 5 days at 20 C. The blank showed negligible depletion. Calculate BOD5.

Working: Depletion = D1 - D2 = 8.6 - 4.1 = 4.5 mg/L. Dilution factor = 100/2 = 50. BOD5 = 4.5 x 50 = 225 mg/L, typical of untreated domestic sewage.

Safety Precautions & Hazard Control

  • Keep incubation strictly at 20 +/- 1 C in complete darkness.
  • Ensure no air bubbles are trapped when stoppering BOD bottles.
  • Use a valid dilution showing 40-70% DO depletion.
  • Do not use samples containing toxic substances without acclimatised seed.

Waste Disposal Protocol

Incubated samples contain sewage microorganisms; disinfect with bleach before disposal to the drain.

Viva Voce Oral Defense Questions & Answers

Q: Why is BOD measured over 5 days at 20 C?

A: The 5-day period at 20 C is the international convention; about 70% of ultimate carbonaceous BOD is exerted in this time, and 20 C approximates average river temperature.

Q: Why is the test done in the dark?

A: To prevent photosynthetic algae from producing oxygen, which would mask the oxygen consumed by bacteria.

Q: What is seeding and when is it needed?

A: Adding acclimatised microorganisms to dilution water; needed for sterile, chlorinated or industrial samples lacking a natural microbial population.