Residual Chlorine in Tap Water: Iodometric Method
To determine residual chlorine (mg/L as Cl2) in tap water by iodometry.
Theoretical Foundations
Cl2 + 2I- -> I2 + 2Cl(-) at pH 4-8; I2 titrated with very dilute thiosulfate (starch). 1 Cl2 = 2 S2O3(2-). The sample is analysed immediately since chlorine escapes and reacts with organics.
Required Apparatus & Chemicals
Laboratory Apparatus
- 50 mL burette
- 250 mL pipette
- 250 mL iodine flask
- Wash bottle
Chemical Reagents
- Tap water (fresh)
- KI crystals
- Acetate buffer (pH 5)
- Standardised Na2S2O3 (0.001000 M)
- Starch indicator
Step-by-Step Practical Procedure
- Pipette 250 mL fresh tap water into an iodine flask.
- Add 1 g KI and acetate buffer.
- Titrate promptly with 0.001000 M thiosulfate, starch near the end.
- Repeat on fresh samples.
- Calculate mg/L as Cl2 (M = 70.90).
Observations & Expected Results
Faint brown I2; blue to colourless endpoint. Typical titre about 4.20 mL.
Worked Stoichiometric Calculations
Problem: 250 mL tap water needed 4.20 mL of 0.001000 M thiosulfate. Find mg/L as Cl2.
Working: Moles thio = 0.001000 x 0.00420 = 0.0000042000 mol. Moles I2 = half = 0.0000021000 mol = moles Cl2. Mass = 0.0000021000 x 70.90 = 0.00014889 g per 250 mL = 0.596 mg/L.
Safety Precautions & Hazard Control
- Analyse immediately; chlorine decays within minutes.
- Use very dilute thiosulfate; levels are sub-mg/L.
- Avoid sunlight during titration.
Waste Disposal Protocol
Dilute iodide solutions may be flushed with water.
Viva Voce Oral Defense Questions & Answers
Q: Why must the analysis be immediate?
A: Residual chlorine is volatile and keeps reacting with dissolved organics/ammonia, so the measured value falls with time.
Q: Why is the titration done near pH 5?
A: Iodide oxidation by chlorine is quantitative there; in strong acid, air oxidation of I- inflates the blank.