CHEMVEN

Dead-Stop Endpoint: Biamperometric Iodometric Titration

To determine Cu2+ iodometrically using a dead-stop (biamperometric) endpoint.

Level: UniversitySafety: Needs supervisionTime: 75 min
Aim of Experiment: To determine Cu2+ iodometrically using a dead-stop (biamperometric) endpoint.

Theoretical Foundations

Two Pt electrodes polarised at ~50 mV: current flows only when the reversible I2/I- couple is present. Before equivalence only I- exists (no current); the first excess I2 depolarises the electrodes and current flows. 2Cu2+ + 4I- -> 2CuI + I2; I2 + 2S2O3(2-) -> 2I- + S4O6(2-). The endpoint needs no starch.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Two Pt electrodes + microammeter
  • 50 mL burette
  • 250 mL beaker
  • Magnetic stirrer
  • 25 mL pipette
  • Small DC source (50 mV)

Chemical Reagents

  • Cu(II) solution (unknown)
  • KI crystals
  • Dilute H2SO4
  • Standardised Na2S2O3 (0.1000 M)

Step-by-Step Practical Procedure

  1. Pipette 25.00 mL Cu solution into a beaker; acidify slightly.
  2. Add 2 g KI.
  3. Insert the Pt electrode pair; apply 50 mV; stir.
  4. Titrate with 0.1000 M thiosulfate.
  5. Endpoint: the microammeter current falls to zero.
  6. Calculate M(Cu2+).

Observations & Expected Results

Current flows during titration, dropping sharply to zero at about 12.40 mL.

Worked Stoichiometric Calculations

Problem: 25.00 mL Cu solution; dead-stop endpoint at 12.40 mL of 0.1000 M thiosulfate. Find M(Cu2+).

Working: Moles thio = 0.1000 x 0.01240 = 0.0012400 mol. Moles I2 = half = 0.00062000 mol. Moles Cu2+ = 2 x 0.00062000 = 0.0012400 mol in 25 mL. M = 0.0012400/0.02500 = 0.049600 mol/L.

Safety Precautions & Hazard Control

  • Keep polarising voltage low (~50 mV) to avoid side electrolysis.
  • Clean Pt electrodes for a sharp current drop.
  • Acidify before adding KI.

Waste Disposal Protocol

Copper/iodide waste to heavy-metal aqueous waste.

Viva Voce Oral Defense Questions & Answers

Q: Why does current flow only near/after the endpoint?

A: Current needs the reversible I2/I- couple at both electrodes; with only I- present (before endpoint) the electrodes stay polarised and no current flows.

Q: What is the advantage over starch?

A: It is objective and works in coloured or turbid solutions where the blue starch colour cannot be seen.