Coulometric Titration: Electrogenerated Hydroxide

To determine the acid content of a sample by coulometric titration with electrogenerated OH-.

Level: UniversitySafety: SafeTime: 75 min
Aim of Experiment: To determine the acid content of a sample by coulometric titration with electrogenerated OH-.

Theoretical Foundations

At a Pt cathode: 2H2O + 2e- -> H2 + 2OH-. The OH- generated is the titrant; charge Q = I x t gives moles: n(OH-) = Q/F (F = 96485 C/mol). No standard solution is needed; the endpoint is detected potentiometrically or with an indicator. Current efficiency must be 100%.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Coulometer / constant-current source
  • Pt electrodes
  • 50 mL burette (for sample)
  • pH meter or indicator
  • Stopwatch
  • 25 mL pipette

Chemical Reagents

  • Acid sample (unknown, dilute)
  • KBr supporting electrolyte
  • Phenolphthalein or pH electrode
  • Distilled water

Step-by-Step Practical Procedure

  1. Pipette 25.00 mL acid sample into the cell with KBr electrolyte.
  2. Insert Pt cathode and anode (anode isolated in a salt bridge).
  3. Apply a constant current of 50.00 mA with stirring.
  4. Stop the current at the phenolphthalein pink endpoint; record time.
  5. Compute Q = I x t; moles OH- = Q/F.
  6. Calculate M(acid).

Observations & Expected Results

Pink endpoint appears; typical generation time about 386 s at 50.00 mA.

Worked Stoichiometric Calculations

Problem: 25.00 mL monoprotic acid; 50.00 mA for 386 s to the endpoint. Find M(acid).

Working: Q = 0.05000 A x 386 s = 19.30 C. Moles e- = 19.30/96485 = 0.00020003 mol = moles OH- = moles H+. M = 0.00020003/0.02500 = 0.0080012 mol/L.

Safety Precautions & Hazard Control

  • Calibrate the current source; the result depends on I x t exactly.
  • Isolate the anode; anodic products must not reach the cathode.
  • Stir efficiently for uniform generation.

Waste Disposal Protocol

Dilute bromide solutions may be flushed with water.

Viva Voce Oral Defense Questions & Answers

Q: What is the primary standard in coulometry?

A: The electron itself, via Faraday's constant; no chemical standard solution is required.

Q: Why must current efficiency be 100%?

A: Every coulomb must generate titrant; side reactions (e.g. O2 reduction) would make Q overestimate the acid.