CHEMVEN

Potentiometric Argentometry: Chloride with Silver Electrode

To determine chloride by potentiometric titration with AgNO3, independent of indicators.

Level: UniversitySafety: Needs supervisionTime: 75 min
Aim of Experiment: To determine chloride by potentiometric titration with AgNO3, independent of indicators.

Theoretical Foundations

A silver wire electrode responds to [Ag+] via Nernst: E = E0 + 0.05916 log[Ag+]. As AgCl precipitates, [Ag+] is fixed by Ksp; at equivalence it jumps ~200 mV. The first-derivative peak gives Veq. Works in coloured/turbid samples where Mohr/Fajans fail.

Required Apparatus & Chemicals

Laboratory Apparatus

  • mV meter
  • Ag wire + reference electrodes
  • 50 mL burette
  • 250 mL beaker
  • Magnetic stirrer
  • 50 mL pipette

Chemical Reagents

  • Chloride sample (unknown)
  • Standardised AgNO3 (0.01004 M)
  • Distilled water

Step-by-Step Practical Procedure

  1. Pipette 50.00 mL sample into a beaker.
  2. Insert electrodes; stir.
  3. Add AgNO3 in 0.50 mL steps, recording mV.
  4. Use 0.10 mL steps across the jump.
  5. Plot E and dE/dV vs V; read Veq.
  6. Calculate mg/L Cl.

Observations & Expected Results

Potential jump of about 200 mV near 14.20 mL; derivative peak sharp.

Worked Stoichiometric Calculations

Problem: 50.00 mL sample; derivative peak at 14.20 mL of 0.01004 M AgNO3. Find mg/L Cl (M = 35.45).

Working: Moles Ag+ = 0.01004 x 0.01420 = 0.00014257 mol = moles Cl-. Mass = 0.00014257 x 35.45 = 0.0050540 g per 50 mL = 101 mg/L.

Safety Precautions & Hazard Control

  • Polish the Ag electrode if the jump is rounded.
  • Stir at constant speed.
  • Shield from light.

Waste Disposal Protocol

Silver waste to silver recovery.

Viva Voce Oral Defense Questions & Answers

Q: Why does the potential jump at equivalence?

A: [Ag+] is pinned low by excess Cl- before equivalence and rises sharply once Ag+ is in excess, per the Nernst equation.

Q: Name an advantage over Mohr's method.

A: No indicator or colour judgment is needed, so turbid, coloured or very dilute samples can be titrated accurately.

← Back to Experiments Directory