Sulfate in Water: Barium-EDTA Back Titration
To determine sulfate in water by precipitating BaSO4 and back-titrating excess Ba2+ with EDTA.
Theoretical Foundations
Excess standard BaCl2 precipitates sulfate: Ba2+ + SO4(2-) -> BaSO4(s). Without filtering, excess Ba2+ is titrated with EDTA at pH 10 (EBT): Ba2+ + H2Y(2-) -> BaY(2-) + 2H+. Sulfate = Ba(total) - Ba(excess). Mg-EDTA is sometimes added to sharpen the endpoint.
Required Apparatus & Chemicals
Laboratory Apparatus
- 50 mL burette
- 100 mL pipette
- 25 mL pipette
- 250 mL conical flask
- Bunsen burner
Chemical Reagents
- Water sample
- Standard BaCl2 (0.01004 M)
- Standardised EDTA (0.01004 M)
- pH 10 buffer
- Eriochrome Black T
- Distilled water
Step-by-Step Practical Procedure
- Pipette 100.0 mL sample into a flask.
- Add 25.00 mL 0.01004 M BaCl2; boil 2 min; cool.
- Add pH 10 buffer and EBT.
- Titrate excess Ba2+ with 0.01004 M EDTA to blue.
- Calculate sulfate as mg/L SO4 (M = 96.06).
Observations & Expected Results
White BaSO4 precipitate; wine-red to blue endpoint. Typical titre about 14.20 mL.
Worked Stoichiometric Calculations
Problem: 100.0 mL sample + 25.00 mL of 0.01004 M BaCl2; back-titre 14.20 mL of 0.01004 M EDTA. Find mg/L SO4.
Working: Total Ba2+ = 0.01004 x 0.02500 = 0.00025100 mol. Excess = 0.01004 x 0.01420 = 0.00014257 mol. Precipitated = 0.00010843 mol = moles SO4(2-). Mass = 0.00010843 x 96.06 = 0.010416 g per 100 mL = 104 mg/L.
Safety Precautions & Hazard Control
- Boil after adding BaCl2 for a filterable precipitate.
- Do not filter; the precipitate does not interfere with the EBT endpoint.
- Add buffer only after cooling.
Waste Disposal Protocol
Barium sulfate is non-toxic but collect barium waste as directed.
Viva Voce Oral Defense Questions & Answers
Q: Why doesn't the BaSO4 precipitate interfere with the titration?
A: BaSO4 is extremely insoluble (Ksp ~ 10^-10); EDTA cannot redissolve it under titration conditions, so only free Ba2+ reacts.
Q: Why is this better than gravimetry for routine water testing?
A: It is much faster (no filtration, drying, weighing) while retaining good accuracy for mg/L levels.