Potentiometric Redox Titration: Fe(II) versus Ce(IV)
To determine Fe(II) by potentiometric titration with Ce(IV) using a platinum electrode.
Theoretical Foundations
Ce4+ + Fe2+ -> Ce3+ + Fe3+. A Pt indicator electrode follows the Nernst equation for the Fe3+/Fe2+ couple before equivalence and the Ce4+/Ce3+ couple after; the potential jumps at equivalence. The first-derivative maximum locates the endpoint without any indicator.
Required Apparatus & Chemicals
Laboratory Apparatus
- pH/mV meter
- Pt + calomel electrodes
- 50 mL burette
- 250 mL beaker
- Magnetic stirrer
- 25 mL pipette
Chemical Reagents
- Fe(II) solution (unknown)
- Standard Ce(SO4)2 (0.1000 M)
- Dilute H2SO4
Step-by-Step Practical Procedure
- Pipette 25.00 mL Fe(II) solution into a beaker; add dilute H2SO4.
- Immerse Pt and reference electrodes.
- Add Ce(IV) in 1.00 mL steps, recording mV after stabilisation.
- Use 0.20 mL steps across the jump.
- Plot E vs V and dE/dV vs V; read Veq at the derivative peak.
- Calculate M(Fe2+).
Observations & Expected Results
Potential rises slowly then jumps about 300 mV within 0.4 mL. Derivative peak at about 21.30 mL.
Worked Stoichiometric Calculations
Problem: 25.00 mL Fe(II); derivative peak at 21.30 mL of 0.1000 M Ce(IV). Find M(Fe2+).
Working: 1:1. Moles Ce4+ = 0.1000 x 0.02130 = 0.0021300 mol = moles Fe2+. M = 0.0021300/0.02500 = 0.085200 mol/L.
Safety Precautions & Hazard Control
- Clean the Pt electrode in nitric acid if the response is sluggish.
- Stir uniformly; wait for stable mV.
- Keep the reference electrode junction clear.
Waste Disposal Protocol
Cerium/iron acidic waste to heavy-metal aqueous waste.
Viva Voce Oral Defense Questions & Answers
Q: Why does the potential jump at equivalence?
A: The electrode switches from the Fe3+/Fe2+ Nernst potential to the much higher Ce4+/Ce3+ potential once Fe2+ is consumed.
Q: Why is Pt used as the indicator electrode?
A: It is inert and conducts electrons reversibly, so it takes up the redox potential of the solution without reacting itself.