Standardisation of Cerium(IV): Sodium Oxalate (Hot)
To standardise Ce(SO4)2 solution against sodium oxalate.
Theoretical Foundations
2Ce4+ + C2O4(2-) -> 2Ce3+ + 2CO2 (hot, H2SO4). Unlike the permanganate standardisation, an indicator is needed: ferroin or N-phenylanthranilic acid. The solution must be hot for a fast reaction.
Required Apparatus & Chemicals
Laboratory Apparatus
- 50 mL burette
- 25 mL pipette
- 250 mL conical flask
- 250 mL volumetric flask
- Thermometer
- Bunsen burner
- Analytical balance
Chemical Reagents
- Ce(SO4)2 solution (approx 0.02 M in H2SO4)
- Sodium oxalate AR
- Dilute H2SO4
- Ferroin indicator
Step-by-Step Practical Procedure
- Weigh about 0.3350 g sodium oxalate; dissolve and make up to 250.0 mL.
- Pipette 25.00 mL; add 10 mL dilute H2SO4; heat to 70 C.
- Add 1 drop ferroin.
- Titrate with Ce(IV) to the pale-blue endpoint.
- Repeat and calculate M(Ce4+).
Observations & Expected Results
Red to pale-blue endpoint. Typical titre about 24.90 mL.
Worked Stoichiometric Calculations
Problem: 0.3350 g Na2C2O4 (M = 134.00) to 250.0 mL; 25.00 mL aliquot needed 24.90 mL Ce(IV). Find M.
Working: Moles oxalate in aliquot = (0.3350/134.00) x 0.1 = 0.00025000 mol. Moles Ce4+ = 2 x 0.00025000 = 0.00050000 mol. M = 0.00050000/0.02490 = 0.02008 mol/L.
Safety Precautions & Hazard Control
- Titrate hot; the cold reaction is sluggish.
- Do not boil; Ce(IV) can oxidise water slowly when very hot.
- Keep the Ce(IV) stock in the dark.
Waste Disposal Protocol
Cerium acidic waste to heavy-metal aqueous waste.
Viva Voce Oral Defense Questions & Answers
Q: Write the half-reaction for oxalate oxidation.
A: C2O4(2-) -> 2CO2 + 2e-.
Q: Why is an indicator needed here but not with permanganate?
A: Ce4+ (yellow) to Ce3+ (colourless) is too faint to see; permanganate's intense purple is self-indicating.