Dissolved Oxygen by Winkler's Method
To determine dissolved oxygen (mg/L) in a water sample by Winkler's iodometric method.
Theoretical Foundations
Mn2+ is precipitated in alkali: Mn2+ + 2OH- -> Mn(OH)2; dissolved O2 oxidises it: 2Mn(OH)2 + O2 -> 2MnO(OH)2. On acidification with I-: MnO(OH)2 + 2I- + 4H+ -> Mn2+ + I2 + 3H2O. 1 O2 = 2 I2 = 4 S2O3(2-). The sample bottle is filled without air bubbles to avoid O2 contamination.
Required Apparatus & Chemicals
Laboratory Apparatus
- 300 mL BOD bottle
- 50 mL burette
- 250 mL iodine flask
- 1 mL and 2 mL pipettes
Chemical Reagents
- Water sample
- MnSO4 solution (48%)
- Alkaline KI solution (KOH + KI)
- Conc. H2SO4
- Standardised Na2S2O3 (0.02500 M)
- Starch indicator
Step-by-Step Practical Procedure
- Fill the BOD bottle completely with sample, avoiding bubbles; stopper.
- Add 1 mL MnSO4 then 2 mL alkaline KI below the surface; stopper without trapping air.
- Invert to mix; let the brown precipitate settle.
- Add 2 mL conc. H2SO4; mix until the precipitate dissolves.
- Titrate 250 mL of the solution with 0.02500 M thiosulfate, starch near the end.
- Calculate DO in mg/L.
Observations & Expected Results
White precipitate turns brown; acidification gives brown I2; starch-blue to colourless. Typical titre about 8.40 mL.
Worked Stoichiometric Calculations
Problem: 250 mL sample needed 8.40 mL of 0.02500 M thiosulfate. Find dissolved O2 in mg/L (M = 32.00).
Working: Moles thio = 0.02500 x 0.00840 = 0.00021000 mol. Moles O2 = 0.00021000/4 = 0.000052500 mol. Mass = 0.000052500 x 32.00 = 0.001680 g per 250 mL = 6.72 mg/L.
Safety Precautions & Hazard Control
- Exclude all air bubbles when stoppering; air adds O2.
- Fix the oxygen immediately at the sampling site.
- Acidify only after the precipitate has settled.
Waste Disposal Protocol
Manganese/iodide waste to heavy-metal aqueous waste.
Viva Voce Oral Defense Questions & Answers
Q: Why is the bottle filled to the brim?
A: Any headspace air would dissolve and oxidise extra Mn(OH)2, giving a falsely high DO.
Q: What interferes with Winkler's method?
A: Nitrite, Fe(III) and organic matter oxidise I-; the azide modification removes nitrite interference.