Standardisation of Potassium Permanganate: Sodium Oxalate
To standardise approximately 0.02 M KMnO4 against sodium oxalate primary standard.
Theoretical Foundations
In acidic solution: 2MnO4(-) + 5C2O4(2-) + 16H+ -> 2Mn2+ + 10CO2 + 8H2O. Sodium oxalate (M = 134.00) is a primary standard. The titration is run at 60-70 C because it is slow at room temperature; Mn2+ product autocatalyses it. KMnO4 is its own indicator (first permanent pink).
Required Apparatus & Chemicals
Laboratory Apparatus
- 50 mL burette
- 25 mL pipette
- 250 mL conical flask
- 250 mL volumetric flask
- Thermometer
- Bunsen burner
- Analytical balance
Chemical Reagents
- KMnO4 solution (approx 0.02 M)
- Sodium oxalate AR (dried)
- Dilute H2SO4 (2 M)
- Distilled water
Step-by-Step Practical Procedure
- Weigh accurately about 1.6750 g sodium oxalate; dissolve and make up to 250.0 mL.
- Pipette 25.00 mL into a flask; add 10 mL 2 M H2SO4.
- Heat to 60-70 C (do not boil).
- Titrate with KMnO4, adding the first drops slowly until the pink decolourises.
- Continue to the first permanent pale pink (30 s).
- Repeat for concordant titres.
Observations & Expected Results
Pink drops decolourise slowly at first, then rapidly; endpoint is a faint pink persisting 30 s. Typical titre about 24.90 mL.
Worked Stoichiometric Calculations
Problem: 1.6750 g Na2C2O4 (M = 134.00) to 250.0 mL; 25.00 mL aliquot needed 24.90 mL KMnO4. Find M(KMnO4).
Working: Moles oxalate in aliquot = (1.6750/134.00) x 0.1 = 0.0012500 mol. Moles MnO4(-) = 0.0012500 x 2/5 = 0.00050000 mol. M = 0.00050000/0.02490 = 0.02008 mol/L.
Safety Precautions & Hazard Control
- Keep the solution hot; cold titration gives a false early endpoint.
- Never use HCl or HNO3 to acidify; HCl is oxidised by MnO4(-).
- Add the first mL very slowly until autocatalysis starts.
Waste Disposal Protocol
Acidic Mn2+ waste: neutralise and collect in heavy-metal aqueous waste.
Viva Voce Oral Defense Questions & Answers
Q: Why is the titration done hot?
A: The oxalate-permanganate reaction is kinetically slow at room temperature; heating to 60-70 C makes it fast and the endpoint sharp.
Q: Why is H2SO4 used and not HCl?
A: Permanganate oxidises Cl- to Cl2 in hot acid, consuming extra MnO4(-) and giving a high result; sulfate is not oxidised.