Spectrophotometric Determination of Trace Lead(II) in Dust as Dithizone Chelate

To determine the concentration of Lead dithizonate quantitatively by measuring optical absorbance at 520 nm and validating the Beer-Lambert law.

Level: University PracticalSafety: Standard Analytical CautionTime: 90 min
Aim of Experiment: To determine the concentration of Lead dithizonate quantitatively by measuring optical absorbance at 520 nm and validating the Beer-Lambert law.

Theoretical Foundations

Quantitative spectrophotometry is founded upon the Beer-Lambert relationship, A = epsilon · b · c, relating monochromatic light attenuation to absorbing species concentration. At the analytical wavelength 520 nm, the analyte Lead dithizonate demonstrates a high molar absorptivity and minimal spectral interference, permitting precise determination down to trace levels.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV-Visible recording spectrophotometer
  • Matched 1.00 cm quartz cuvettes
  • Precision volumetric flasks (50 mL, 100 mL)
  • Calibrated micropipettes
  • Analytical balance (±0.0001 g)

Chemical Reagents

  • Standard reference stock of Lead dithizonate
  • Specific chromogenic / complexing reagent solutions
  • Analytical buffer / matrix modifiers
  • Analytical reagent blank solutions
  • Ultra-pure deionised water

Step-by-Step Practical Procedure

  1. Prepare a series of five standard working solutions of Lead dithizonate spanning the linear absorbance range (0.1 to 0.9 AU).
  2. Prepare the unknown sample solution according to the stoichiometric protocol.
  3. Configure the spectrophotometer to the analytical wavelength 520 nm.
  4. Zero the instrument absorbance against the matrix reagent blank in a 1.00 cm cuvette.
  5. Record absorbance values in triplicate for all calibration standards in ascending concentration order.
  6. Measure the absorbance of the unknown sample under identical optical parameters.
  7. Perform linear regression analysis (A vs C) to determine sensitivity slope (m) and correlation coefficient (R^2 > 0.999).
  8. Calculate the exact analyte concentration in the unknown sample accounting for all dilution factors.

Observations & Expected Results

Solutions exhibit characteristic chromophore coloration. Absorbance exhibits strict linearity with concentration across the working range at 520 nm.

Worked Stoichiometric Calculations

Problem: Given calibration slope m = 0.0520 L/mg and unknown absorbance A = 0.416 at 520 nm, calculate the concentration of Lead dithizonate.

Working: Concentration C = A / m = 0.416 / 0.0520 = 8.00 mg/L Lead dithizonate. Account for any pre-dilution factors to report final sample concentration.

Safety Precautions & Hazard Control

  • Carefully clean optical faces of cuvettes with lens paper before insertion into the beam.
  • Zero the instrument with an authentic reagent blank containing all reagents except the analyte.

Waste Disposal Protocol

Collect all chemical solutions in dedicated waste receptacles according to laboratory safety protocols.

Viva Voce Oral Defense Questions & Answers

Q: Why is 520 nm chosen as the measurement wavelength?

A: At 520 nm (lambda_max), the molar absorptivity is maximized, providing the greatest sensitivity (Delta A / Delta C) and minimizing deviations from Beer's law caused by polychromatic light.

Q: What causes chemical deviations from Beer's law in spectrophotometry?

A: Chemical deviations occur when the absorbing species participates in concentration-dependent equilibria such as association, dissociation, polymerisation, or reaction with the solvent or buffer.